AM852 Veterinary Anaesthesia machine User’ Manual
12
Operation
Storage
Temperature
+10
℃~
+40
℃
-5
℃~
+50
℃
Relative
humidity
30%
~
75% no
condensation
10%
~
90% no
condensation
Pressure
70 kPa
~
106kPa
50 kPa
~
106kPa
Anaesthetic
concentration
0.5%
~
5.0%
The carrier
gas(O
2
N2O)
flow
0.2L/min
~
10L/min
2.5Instructions for filling the AVDD
1. Check that the vaporizer concentration control is in the ‘0’ position.
2. Unscrew the filler cap.
3. Remove the bottle cap and fill the vaporizer slowly and carefully, stopping to check
the liquid level occasionally, Stop filling when the upper mark is reached on the
filler block.
4. Check that the seal in the filler cap is clean, and positioned correctly. Replace the
filler cap. Tighten finger tight only. DO NOT use a wrench.
Capacity
Volume at MAX mark
150ml(nominal)
Volume at MIN mark
35ml(nominal)
Determination of vapour concentration delivered by AVDD
1. Test the
AVDD
on a calibrated test rig capable of supplying the necessary gas flowrate
and pressures required by the test conditions, or on an anaesthetic system with the
anaesthetic ventilator and anaesthetic breathing system supplied or recommended by
the manufacturer or supplier.
Connect an anaesthetic vapour analyser to the fresh gas outlet of the anaesthetic
system, or to the inlet of the anaesthetic breathing system if there is no fresh gas outlet
or, if applicable, to the inspiratory port of the anaesthetic ventilator.
Check to ensure that the components downstream of the
AVDD
will not affect the test
results, for example by absorbing volatile agents, by imposing time delays on
response, or by leakage.
2. Condition the calibrated test rig or the anaesthetic system, as applicable, with the
specified test equipment and anaesthetic agent in the test room for at least 3 h