AM852 Veterinary Anaesthesia machine User’ Manual

27

8.Trouble shooting

Warning:

Please make eligible service representative repair the system.

Fault symptom and

alarm info

Possible cause

Troubleshooting

The animal respiratory

circuit leak gas

APL valve disclose by manual
control

Close APL valve

Absorber fix untightly

Fix it again

Corrugated hose mangle or joint
loose

Replace hose or fix again

Inhale or exhale unidirectional
valve cover doesn’t screw down

Screw again

Manual switch doesn’t work

Please contact with our company

Airway pressure is far

high when manual

respiration

APL valve adjust isn’t accurate

Adjust APL valve to right
position

دﯾﯾﺎﻣرﻓ ﮫﻌﺟارﻣ.WWW.ELM-FAN.IRتﯾﺎﺳ ﮫﺑ رﺗﺷﯾﺑ دراوﻣ دوﻠﻧاد یارﺑ

To download more items, go to ELM-FAN.IR

Summary of Contents for AM852

Page 1: ...ation 19 4 6Test Respiration system 20 5 Installation and Replacing 22 5 1Install absorber 22 5 2 Replace absorbent 22 6 Cleaning and Sterilization 23 6 1Cleaning and sterilization of Anaesthesia work...

Page 2: ...achine User Manual 3 9 1Diagram of system drive air circuit 28 10 Technical Specification 30 11 Detailed list of host machine accessories and consumer goods 34 12 Guarantee 35 WWW ELM FAN IR To downlo...

Page 3: ...carbon dioxide exhaled by the animal in an operation 1 2Symbols used in the manual and device Warning and Note show the possible risks if you disobey the declarations in this manual Please read and ab...

Page 4: ...embly in this device and please refer to the local standards for the relevant rules of detection Importance Punchy recommendation When using the system User should equip anaesthetic gas spiritualizati...

Page 5: ...ns in the animal s body React with Anaesthesia gas or anesthetic to produce dangerous byproducts Warning The danger of explosion Ensure not to use flammable anesthetic in AM852 system Warning Do not u...

Page 6: ...sthetic system 2 1Anaesthetic control system 1 O2 Flush press button 2 Flowrate regulating 3 Evaporator 4 Respiration system 5 NO2 pressure gauge 6 O2 pressure gauge Figure2 1 AM852 Control functions...

Page 7: ...owrate O2 to respiration system 2 Flowrate regulating Revolving control button anti clockwise will augment flowrate and revolving control button clockwise will decrease flowrate 1 Pipeline gas linker...

Page 8: ...piration system Items Remark absorber load and unload Unscrewing screw releasing absorber and then can unload it fitting on absorber and screw screw can load it APL valve Adjustable pressure limit val...

Page 9: ...d Connect exhaust outlet of respiration system with AGSS by exhaust emission tubes according with EN740 standard See Fig 2 4 1 Respiration system 2 Connecting tube 3 AGSS system Fig 2 4 Connecting AGS...

Page 10: ...marks 1 Drug adding knob Anticlockwise rotate screw cap to unload it and the drug can be put in clockwise rotate screw cap to airproof 2 Lock knob Press this button and rotate concentration control sw...

Page 11: ...minal Determination of vapour concentration delivered by AVDD 1 Test the AVDD on a calibrated test rig capable of supplying the necessary gas flowrate and pressures required by the test conditions or...

Page 12: ...of the fresh gas outlet pressure at the end of the inspiration period to 33 of this pressure is less than 0 6 s Note This can be achieved by using a test lung having a compliance of 0 2 l kPa and an...

Page 13: ...sure the output of anaesthetic vapour concentration of vapour volume of gas for 1 min before during a 10s activation of the oxygen flush and for 30s after the oxygen flush activation Compare these thr...

Page 14: ...machine User Manual 15 3 Assembly of whole set 3 1Respiratory system installation 1 Common gas pipe 2 Holder 3 Respiration system Fig 3 1 Pipeline installation Fig 3 2 1 2 3 WWW ELM FAN IR To download...

Page 15: ...stem applicable Test pipe and gas cylinder applicable Test flowrate control applicable Install evaporator applicable Test respiration system applicable applicable Warning Do not put the machine into o...

Page 16: ...lt in storing supply shortage when pipeline is in trouble 7 The contingency equipment have already been prepared and in good condition 8 The equipment used to maintain airway tracheal intubatton and m...

Page 17: ...eligible method to collect and dispose it Unsuitable mixture gas will harm the animal If the linkage system can not supply the suitable proportioned O2 and N2O the system should not be used 1 Connect...

Page 18: ...stop at last 8 Clockwise rotate all flowrate regulating switches ensure to rotate them to the bottom 9 Connect the pipes to the system again 4 5Evaporator Installation Warning The evaporators which c...

Page 19: ...and descends during the expiration period Inspiration unilateral valve 2 ascends during the inspiration period and descends during the expiration period Figure 4 1 Respiration System 3 Gas leak test o...

Page 20: ...the value of airway pressure do not over 7 5kPa Certain pressure fluctuation is normal circumstance Open the APL valve completely Ensure the value of airway pressure do not over 0 5kPa Press the O2 fl...

Page 21: ...r change of calx sodica in absorber indicates CO2 absorption but the color change of calx sodica is a approximate show Utilize a CO2 detector to determine whether to replace absorbent or not Discard t...

Page 22: ...bber corrugated tube and manual breathing bag Every used rubber corrugated tube and manual breathing bag must be firstly washed with rinsing carefully and after they were dried put them into the suffo...

Page 23: ...te in accordance with the required regulation and checking procedure Please handle all parts with care lest valve cover and cover plate are bumped or crushed 6 4Cleaning and sterilization of absorber...

Page 24: ...ion Check whether accessories are in good condition or not and replace or service them when necessary Process according to need Empty water store and replace absorbent in absorber 7 2Maintenance of an...

Page 25: ...urther rotate knobs forcibly when reaching already the max or min otherwise they would be damaged and malfunction 7 4Maintenance of inspiration and expiratory unidirectional valve Do not press the cov...

Page 26: ...disclose by manual control Close APL valve Absorber fix untightly Fix it again Corrugated hose mangle or joint loose Replace hose or fix again Inhale or exhale unidirectional valve cover doesn t screw...

Page 27: ...2 Veterinary Anaesthesia machine User Manual 28 9 Operation Principle 9 1Diagram of system drive air circuit Figure 9 1 diagram of drive air circuit WWW ELM FAN IR To download more items go to ELM FAN...

Page 28: ...O2 pressure until N2O has been cut off Ensure the concentration of O2 at fresh gas outlet is more than 25 When the pressure of O2 is too low alarm device sends sound of low O2 alarm Mixture Gas From t...

Page 29: ...pipeline inlet 280kPa 600kPa Note Gas providing must be medicinal level Note When the medical gas pipeline supply pressure is lower than 0 15Mpa the anaesthetic system will cease to deliver gas Flowme...

Page 30: ...hetic breathing system shall not exceed 150 ml min 15 2 kPa l min at 3 0 kPa 30 cm H2O internal pressure Compliance The internal compliance of the anaesthetic breathing system shall not exceed 120 ml...

Page 31: ...ading is between 1 and 5cm H2O 8 For semi closed circuits check the continuity of the inner tube by passing a flow of 20l m along the inner tube with the APL valve closed Block the outlet for a time s...

Page 32: ...AM852in vibrate environment System Height 450mm Width 405mm Depth 330mm Weight 21 25kg Environment Requirement Operation Storage Temperature 15 35 5 40 Relative humidity 90 no condensation 90 no cond...

Page 33: ...Spare part evaporator 1 Spare part Respiratory circuit system 1 Spare part accessories O2 feeding pipeline 1 Spare part N2O feeding pipeline 1 Spare part expendables Breathing pipeline 1 set Spare pa...

Page 34: ...e assemblies is 30 days The above assurances are invalid in the following status the repair works are accomplished by other companies the repair works are not carried out according to the written inst...

Page 35: ...esia machine User Manual 36 Operation Manual for AM852 Veterinary Anaesthesia machine Version 15 01 REVISED IN January 2015 Beijing Eternity Electronic Technology CO LTD WWW ELM FAN IR To download mor...

Reviews: