Pre-use Test
4.1Test interval
Under the following condition
,
pre-use test should carry out:
1. Before the first pet use every day.
2. Before each pet use.
3. Test machine according to the request after service and maintenance.
Test item
Before the first
pet use every day
Before each pet use
Check system
applicable
-
Test pipe and gas cylinder
applicable
-
Test flowrate control
applicable
-
Install evaporator
applicable
-
Test respiration system
applicable
applicable
Warning:
Do not put the machine into operation before you read and comprehend
the operation and maintenance of each parts.
·
All system connection.
·
All warnings and notes.
·
Operation method of each parts.
·
Test method of each parts.
Before you use the system, should
·
Complete all tests mentioned in this section.
·
Test all other system parts.