30
Table 3: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity – for equipment and
systems that are not life‐supporting
ELITONE is intended for use in the electromagnetic environment specified below. The customer or the user
should assure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Electromagnetic Environment ‐ Guidance
Radiated RF
IEC 61000‐4‐3
10 V/m
80 MHz to
2.7 GHz
10 V/m
80 MHz to
2.7 GHz
Portable and mobile RF communications equipment should
be used no closer to any part of ELITONE, including cables,
than the recommended separation distance calculated from
the equation application to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.7 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) and
d
is the recommended
separation distance in meters (m).