29
Table 1: Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
ELITONE is intended for use in the electromagnetic environment specified below. The customer or the user
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment ‐ Guidance
RF emissions CISPR 11
Group 1
ELITONE uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
ELITONE is suitable for use in all establishments including
domestic and those directly connected to the public supply
network that supplies buildings used for domestic purposes.
Table 2: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity – for all equipment
ELITONE is intended for use in the electromagnetic environment specified below. The customer or the user
should assure that it is used in such an environment.
Immunity Test
Test Level
Compliance
Electromagnetic Environment ‐ Guidance
Electrostatic Discharge
(ESD) IEC 61000‐4‐2
± 8kV Contact
± 15kV Air
± 8kV Contact
± 15kV Air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.