DELAY TIME IN AC MODE
5.
POWER SUPPLY CONNECTION
The output of RPS is not isolated respect to the main power supply therefore NONE
OF THE TWO OUTPUT PHASES CAN BE CONNECTED TO THE SAFETY GROUND.
5.1.
FUSES
Fuses can be used to protect the power line of the RPS. We recommend to use a delayed fuses
with rated current of 16 A
RMS
and a rated voltage of 230 V
RMS
(Both for phase and neutral).
5.2.
RCD PROTECTION
Below the recommended RCD protections (Residual Current Device) for the input line. It's
recommended to use
B type
RCD with a earth leakage current of
30 mA
.
5.3.
MAGNETO THERMAL PROTECTION
The Magneto-thermal circuit breaker protect the input line from short circuits. Generally depends
on the load and on the connection (section and length of the cable). It is recommended to use a
magneto-thermal protection with
type C
curve.