Page 31 of 40 M8144
Issue 2.0
8.
TECHNICAL SPECIFICATION.
8.1. The specified data for each instrument is shown in the table below.
IA9100
IA9200
IA9300
32- bit processor.
3 capillary Tubes up 2mm OD (storage for pack).
Tube guide removal for cleaning and use of cold finger
40mm dia lens (magnification x 8) removable and with
ajustable focus.
13
– key membrane keypad
Sample presevation ceiling temperature selection
White
3 x Super Flux LED’s
Selectable ramp rate of 1.0°C/min and 10.0°C/min
Ramp rate Selectable with a 1.0°C/min default setting
ranging from 0.2°C/min to 10°C/min.
Date and time facility
PC output facility
Printer output facility
USB output to Pen key.
4 x 1 Melt memory capacity
3 x 2 Melt memory capacity
Batch memory capacity of 500 melts
Batch memory capacity of 2000 melts
Serial printer connection facility
Pack of 50 capillary tubes
8.2. General specification.
Electrical specification
Total power consumption
45W
Oven control
Closed loop P.I.D.
Oven illumination
3 x white Super Flux LED’s
Illuminous Intensity (per LED)
3500 Lux
Oven operating temperature range.
5°C to 400°C
Power supply PSU
When ordering a replacement PSU, always quote the IA9000 instrument’s serial number.
Input voltage
100
– 240V~50.60Hz.
Output voltage
15V DC
General specification
Temperature display range
1°C to 405°C
Display resolution
0.1°C
Accuracy
± 0.5°C @ 20°C linerly to ± 1.0°C @ 350°C
Ramp rate
From 0.2°C/min to 10°C/min settable in increments
of 0.1°/min (Not 9100)
Ceiling temperature setting
From 5°C above set temperature to 405°C in
increments of 1°C
Time and date
Settable with date display as MMDDYY or
DDMMYY. Time selectable in 12 or 24 hour format.
(Not 9100)
Display
12 x 2 backlit matrix LCD
Control Sensor
Platinum Resistance Probe (PT100).