INSTALLATION
12/2021
17
NP-DO M-39_12-2021
INSTALLATION
Risk of incorrect installation.
Only a qualified professional can install the equipment and place it into operation
for the first time. This professional is obliged to train operating staff as to the use
and maintenance of the device. Installation and training of all operators shall be
confirmed by the installer’s signature on the certificate of installation.
Risk of incorrect installation.
ANY MODIFICATION OF THIS EQUIPMENT IS FORBIDDEN!
Risk of damage to the device.
This equipment cannot be used nearby other instruments. If this equipment is
used nearby other instruments, the equipment must be observed in order to verify
normal operations in the configuration it will be used.
Instruments may be affected electro-magnetically!
You may notice a “new product” odour when you first place the product into
service (for a short period of time). This odour is temporary and does not impede
the normal use of the product. Ensure the space is properly ventilated after
installation
.
INSTALLATION CONDITIONS
The equipment may only be installed and operated in dry, well ventilated and dust-free rooms.
The mobile dental suction equipment should be installed with regard to an easy access for the
operation and maintenance and good accessible rating plate.
The equipment should stand on a plane, sufficiently stable base (be aware of the suction
equipment weight, see Technical data).
The equipment may not be operated outdoors or in wet or damp environments.
Risk of explosion.
Do not use the equipment in the presence of explosive gases, dust or
combustible liquids.
To guarantee a trouble-free operation of the suction equipment, the room temperature should
never drop under +5˚C and 40°C. The ideal ambient temperatures are b10°C
up to +25°C.
Most of the electric energy used by the suction pump (9) is converted to heat and given off to the
ambient. During a longer lasting work, above all with the desalivating mouthpiece, the box
temperature increases over 40°C causing an automatic actuation of the cooling fan. After the
room is cooled down under approximately 32°C, the fan turns off again.
Summary of Contents for ASPINA DO M
Page 2: ......
Page 4: ...CONTENTS 5 INHALT 31 TABLES DES MATI RES 57 83 SPIS TRE CI 110 OBSAH 136 OBSAH 162...
Page 84: ...NP DO M 39_12 2021 84 12 2021 93 42 EEC MDD DO M CE...
Page 85: ...12 2021 85 NP DO M 39_12 2021 B 5 1...
Page 86: ...NP DO M 39_12 2021 86 12 2021 5 2 5 3...
Page 87: ...12 2021 87 NP DO M 39_12 2021...
Page 88: ...NP DO M 39_12 2021 88 12 2021 6 1 25 55 C 24 70 C 90...
Page 90: ...NP DO M 39_12 2021 90 12 2021 1 ASPINA DO M...
Page 91: ...12 2021 91 NP DO M 39_12 2021 1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18...
Page 96: ...NP DO M 39_12 2021 96 12 2021 5 C 40 C 10 C 25 C 40 C 32 C...
Page 97: ...12 2021 97 NP DO M 39_12 2021 230 11 10 13 5 I 3 11 16 14 8...
Page 98: ...NP DO M 39_12 2021 98 12 2021 12 1 Aspina DO M 1 2 3 4 5 6 7 8...
Page 100: ...NP DO M 39_12 2021 100 12 2021 5 I 6 1 2 11 6 10 9 4 10 8 1 12 13 1 17 7 5 O...
Page 101: ...12 2021 101 NP DO M 39_12 2021 13 2 13 1 1 2 7 14 8 14 8 100 B IEC 60601 1...
Page 102: ...NP DO M 39_12 2021 102 12 2021 24 EN 62353 EN 62353 G CE...
Page 104: ...NP DO M 39_12 2021 104 12 2021 14 2 7 1 14 1 6 2 5 O 14 3 1 Lubri Jet 15 14 4 11 1 14 1 6 14 5...
Page 107: ...12 2021 107 NP DO M 39_12 2021 14 9 1 2 3 2 14 2 14 4 I 15 20 O 14 1...
Page 108: ...NP DO M 39_12 2021 108 12 2021 14 I o 2 T 6 3 A...
Page 109: ...12 2021 109 NP DO M 39_12 2021 14 1 14 1...
Page 116: ...OPIS PRODUKTU NP DO M 39_12 2021 116 12 2021 Rys 1 Mobilna odsysarka dentalna ASPINA DO M...
Page 142: ...POPIS V ROBKU NP DO M 39_12 2021 142 12 2021 Obr 1 Mobiln dent lna ods va ka ASPINA DO M...
Page 168: ...POPIS V ROBKU NP DO M 39_12 2021 168 12 2021 Obr 1 Mobiln dent ln ods va ka ASPINA DO M...
Page 195: ......