EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4
Pg. 41
3. Data conversion when the data format is FLOAT [32 base numbers]
- Output data format: FLOAT [hexadecimal number]
- Converted data format: decimal number
Conversion Example: Data conversion of solar irradiance [Address 2 to 3]
The FLOAT data will be in the following order:
As shown in the example, put the data for No. 2 [before] and the data for No. 3 [after] together to
make them as one data before conversion.
Next, calculate the data before conversion according to the IEEE754 standard.
Conversion result: 820.52 [W/m2]
Address No.
2
3
Output Data
21
47
44
4d
Pre-conversion
data
444D2147
Conversion result
820.52