Displayed contents [configuration] of Main Window are shown below.
Fig. A-5. Main Screen
Table A-3. Main screen display contents
#
Notation name
Details
1
Modbus Addr.
Modbus address number [1 to 100] to be connected
[Default: The last 2 digits of the MS-60S serial number is setup as address
number. If the number is
“0X”, the “X” would be the address number. In the
case of
“00”, the address would be “100”.
2
Comm.Setting
Communication setting. This can be selected from the combination of baud rate
[2400/4800/9600/19200/38400/115200] and parity check [None/Even/Odd].
[Default: 19200/None]
3
Registor Type
Registor type can be selected from 4 types: EKO [Default]/MC-20/KZ Type/HF
Type.
*Use only when communication installation of master device does not change in
the case of installing different product, such as swapping pyranometers.
Registry responses corresponding against the requests to registry number for the
following product before swapping:
MC-20: compatible to MS-80M/60M/40M
KZ Type: compatible to Kipp&Zonen SMP-3/6/10/11/21/22
FH Type: compatible to Hukseflux
SR30-D1/SR20-D2/SR15-D1/SR15D-D2A2/SR05-D1A3/SR05-D1A3-PV/
SR05-D2A2
4
Firmware Ver.
Firmware version information [cannot be changed]
5
Hardware Ver.
Hardware version information [cannot be changed]
6
MFG. Date
Manufacturing date information [cannot be changed]
7
S/N
Product serial number [cannot be changed]
8
Name
Product name [cannot be changed]
9
Calibration Date/
Product calibration date, sensitivity [cannot be changed]