Table A-9: Communication setting
#
Communication
speed [bps]
Parity bit
#
Communication
speed [bps]
Parity bit
0
2400
None
10
19200
Even
1
2400
Even
11
19200
Odd
2
2400
Odd
12
38400
None
3
4800
None
13
38400
Even
4
4800
Even
14
38400
Odd
5
4800
Odd
15
115200
None
6
9600
None
16
115200
Even
7
9600
Even
17
115200
Odd
8
9600
Odd
9
19200
None
K.
Register Type [Address Number: 105]
Select from 4 modes to operate. When operated with other than the mode 1, some functions will be
unavailable. {1: Original, 2: MS-XXM mode, 3: KZ mode, 4: HF mode}
L.
Digital Filter Constant [Address Number: 106] Value: 1000 [Fixed]
M.
Analog Input Setting [Address Number: 107]
From the combination of amplifier type and input range, it is selected with number [#].
Do not change from the factory setting.
TableA-10: Analog input setting
#
Amplifier type
Input range
#
Amplifier type
Input range
0
* OPA
200 mV
5
PGA
256 mV
1
OPA
20 mV
6
PGA
128 mV
2
** PGA
2048 mV
7
PGA
64 mV
3
PGA
1024 mV
8
PGA
32 mV
4
PGA
512 mV
9
PGA
16 mV
* OPA: Operational amplifier
、
** PGA: Programmed gain amplifier
N.
Analog Output Setting [Address Number: 108] Value: 0 [Default]
Select and setup from the following numbers: {0: No analog output, 1: 0-1V output, 2: 4-20mA, 3: Manual}
O.
Pt100 Calibration Value [Address Number: 129, 130]
Pt100 offset can be corrected by ratio [
“1” = 100%, no correction]
P.
Analog Output Calibration Value [Address Number: 131, 132]
Correction value for analog output power
Q.
Load Resistor for 0-1V Output [Address Number: 133, 134] Value: 100 [Default]
Precision resistor value [
Ω] used when analog 0-1V output is selected.