EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4
Pg. 37
192, 193
UINT32
Calibration history 2[Calibration date,YYYYMMDD
type]
194, 195
FLOAT32
Calibration history 2[Sensitivity [
μ
V/W/m
2
]]
196, 197
UINT32
Calibration history 3[Calibration date,YYYYMMDD
type]
198, 199
FLOAT32
Calibration history 3[Sensitivity [
μ
V/W/m
2
]]
200, 201
UINT32
Calibration history 4[Calibration date,YYYYMMDD
type]
202, 203
FLOAT32
Calibration history 4[Sensitivity [
μ
V/W/m
2
]]
[Detailed description of retained data]
A.
Model Number [Address Number: 0]
Pyranometer: ”0x0110”, Pyrgeometer: “0x0120” *Do not change from the factory setting.
B.
Status [Address Number: 1] Value: 0 [
Always “0”]
C.
Tilt angle, Role angle [Address Number: 80, 81 and 82, 83] Unit: °
Tilt angle: North-South tilt angle. When the pyranometer cable is facing north, the inclination to the west is
indicated with positive value and the inclination to the east is indicated with minus. Maximum angle: ± 90 °
Role angle: East-West tilt angle. When the pyranometer cable is facing north, the inclination to the west is
indicated with positive value and the inclination to the east is indicated with minus. Maximum angle: ± 90 °
D.
X-axis/Y-axis/Z-axis Counts [Address Number: 84 to 86]
Values corresponding to the tilt for each axis of tilt sensor [±2G =±8,192].
E.
Message Count Value [Address Number: 90]
Data counts received properly.
F.
CRC Error Count Value [Address Number: 91]
Data counts failed to acquire with CRC error.
G.
Exception Error Occurrence Count Value [Address Number: 92]
Error counts failed to acquire due to exception error caused by some abnormality.
H.
Product Type [Address Number: 102]
Apply the arithmetic equation for each product. Do not change from the factory setting .
I.
Modbus Address [Address Number: 103]
The last 2 digits of the product serial number is setup as address number in default.
Ex.] S/N SXXXXXX05: Address = 5, S/N SXXXXXX00: Address = 100 [
when “00”, the address is ”100”]
J.
Serial Communication Setup [Address Number: 104] Value: 6 [Default]
Select with number [3] from the combination of baud rate and parity bit.