EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4
Pg. 17
2-5. Wire Assignments
Also see [7-3. Output Cables].
Table 5-3. Wire Color Codes [MS-60]
Table 5-4. Color Codes of cable[MS-60S]
No.
Cable Color
4-20mA
Modbus
SDI-12
0-1V *
1.
Brown
DC 5 to 36V [+]
DC 5 to 36V [+]
DC12V
DC5 to 36V [+]
2.
White
4-20mA [-]
/ GND
GND
GND
0-10mA[-]
/ 0-1V[-] / GND
3.
Blue
---
RS485/B/+
SDI-12 Data
---
4.
Black
---
RS485/A/
-
---
---
5.
Gray
4-20mA [+]
---
---
0-10mA[+]
/ 0-1V[+]
Shield
Shield
FG
FG
FG
FG
*When selecting 0-1V output, a precision resistor is required separately. The output depends on the
accuracy of the resistor.
No.
Cable Color
MS-60
1.
White
mV [+]
2.
Black
mV [-]
Shield
Shield
FG
Figure 5-6. Connector pin number of MS-80/80S
E
ach number corresponds to the number in Table 5-4.