EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4
Pg. 35
Table A-8. Retained data contents in the registry
Address number: Only 0 to 99 Read
、
Read/Wire is possible after 100
Address
Data Type
Description
Details
[described later]
0
UINT16
Model Number
A
1
UINT16
Status
B
2, 3
FLOAT32
Solar Irradiance [W/m
2
]
50, 51
FLOAT32
AD Input voltage [mV]
52, 53
FLOAT32
Pt100 thermistor value [
Ω
]
Measurement range
{
0 to 250
Ω}
54, 55
FLOAT32
Pt100 temperature [
℃
]
56, 57
FLOAT32
Internal temperature [
℃
]
58, 59
FLOAT32
Internal humidity [RH%]
60
UINT16
Power supply voltage value [mV]
61
UINT16
Reference voltage value [=2,048 [mV]]
62
UINT16
Digital
→
Analog conversion voltage [mV]
63
UINT16
Analog power supply voltage [5,000 [mV]]
64
UINT16
Digital power supply voltage [3,300 [mV]]
80, 81
82, 83
FLOAT32
FLOAT32
Tilt Angle [
°
]
Role Angle [
°
]
C
84
85
86
SINT16
X axis count value
Y axis count value
Z axis count value
D
90
UINT16
Message count value
E
91
UINT16
CRC Error count value
F
92
UINT16
Exception error occurrence count value
G
100
UINT16
Firmware version
101
UINT16
Hardware version
102
UINT16
Product type [Switching between
Pyranometer/Pyrgeometer/UV radiometer]
H
103
UINT16
Modbus address
I
104
UINT16
Serial communication setting
J
105
UINT16
Register type
K
106
UINT16
Digital filter time constant
L
107
UINT16
Analog input setting
M
108
UINT16
Analog output setting
N
109, 110
to
127, 128
FLOAT32
Calibration value of analog input setting
9 types in combination of amplifier type and range