38
Appendix
EMC Information
The DuraVision FDS1782T
has a performance that appropriately displays images.
Environments of Intended Use
This product is intended for use in Professional healthcare facility environments such as clinics and
hospitals. The following environments are not suitable for the RadiForce series to be used:
• Home healthcare environments
• In the vicinity of high-frequency surgical equipments such as electrosurgical knives
• In the vicinity of short-wave therapy equipments
• RF shielded room of the medical equipment systems for MRI
• In shielded location Special environments
• Installed in vehicles including ambulances.
• Other special environments
WARNING
The DuraVision FDS1782T requires special precautions regarding EMC and need to be installed.
You need to carefully read EMC Information and the "PRECAUTIONS" section in this document, and
observe the following instructions when installing and operating the product.
Be sure to use the cables attached to the product, or cables specified by EIZO.
Use of cables other than those specified or provided by EIZO of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this equipment
and improper operation.
The DuraVision FDS1782T should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the DuraVision FDS1782T or system should be observed to
verify normal operation in the configuration in which it will be used.
When using a portable RF communication equipment, keep it 30 cm (12 inches) or more away from
any part, including cables, of the DuraVision FDS1782T. Otherwise, degradation of the performance
of the DuraVision FDS1782T could result.
Anyone who connects additional equipment to the signal input part or signal output parts,
configuring a medical system, is responsible that the system complies with the requirements of
IEC/EN60601-1-2.
Technical Descriptions
Electromagnetic emissions
The DuraVision FDS1782T is intended for use in the electromagnetic environment specified below.
The customer or the user of
the DuraVision FDS1782T
should assure that
the DuraVision FDS1782T
is used in the following
environment.
Emission test
Compliance
Electromagnetic environment - Guidance
RF emissions
CISPR11 / EN55011
Group 1
The DuraVision FDS1782T
uses RF energy only for its internal functions.
Therefore, its RF emission are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11 / EN55011
Class B
The DuraVision FDS1782T
is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC / EN61000-3-2
Class D
Voltage fluctuations /
flicker emissions
IEC / EN61000-3-3
Complies