41
Appendix
Recommended separation distances between portable or mobile RF communication equipment,
and recognized communication equipment services
The DuraVision FDS1782T
is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the
DuraVision FDS1782T
can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
DuraVision FDS1782T. Immunity to proximity fields from following RF wireless communication equipments has been
confirmed:
Test
frequency
(MHz)
Bandwidth
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Minimum
separation
distance
(m)
IEC/EN60601 test
Level (V/m)
Compliance
level
(V/m)
385
380 - 390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
27
450
430 - 470
GMRS 460,
FRS 460
FM
±5 kHz deviation
1 kHz sine
2
0.3
28
28
710
704 - 787
LTE Band 13,17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
9
745
780
810
800 - 960
GSM 800 / 900,
TETRA 800,
iDEN 820
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
28
870
930
1720
1700 - 1990 GSM 1800;
CDMA 1900;
GMS 1900;
DECT;
LTE Band 1, 3, 4,
25;
UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
28
1845
1970
2450
2400 - 2570 Bluetooth,
WLAN,
802.11 b/g/n,
FIRD 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
28
5240
5100 - 5800 WLAN 802.11 a/n Pulse
modulation
b)
217 Hz
0.2
0.3
9
9
5500
5785
a) For some services, only the uplink frequencies are included.
b)
carrier are modulated using a 50 % duty cycle square wave signal.