
Installation & User Manual - Sampling Systems
Version 140618 UK
Page 31
7.4
Installation instructions
•
Mount the sampler water levelled to a vertical wall with the
supplied bracket. Ensure there is enough height for a proper
discharge of the sample to the container.
•
Ensure a wide dimensioning of the flush pipe (minimum 50mm).
Be Aware! Flush for a representative sample min. 3x the volume
of the supply pipe.
•
Mount the supply pipe to a 100% filled pipe (see fig 7.4a). Keep
the supply pipe as short as possible.
Be aware! The supply pipe
needs to be disconnected for safe maintenance and reparations.
•
Place the sample drain hose over the pillar from the sample outlet
of the sampler.
Be aware! Keep the length of the sample drain
hose as possible and free from siphons and nods.
7.5 Changing sample volume
Contact your supplier.
Fig. 7.4b
Fig. 7.4a
Summary of Contents for omy
Page 44: ...Installation User Manual Sampling Systems Version 140618 UK Page 44 ...
Page 45: ...Installation User Manual Sampling Systems Version 140618 UK Page 45 ...
Page 46: ...Installation User Manual Sampling Systems Version 140618 UK Page 46 ...
Page 47: ...Installation User Manual Sampling Systems Version 140618 UK Page 47 ...
Page 48: ...Installation User Manual Sampling Systems Version 140618 UK Page 48 ...