Installation & User Manual - Sampling Systems
Version 140618 UK
Page 16
3.5.3
Changing Sample Settings
Select submenu
Sample Settings
and press
Enter
.
Select from the following parameters. Press
Enter
to change its value.
Parameter
Description
Sample by
Select with
◄
or
►
the following:
Volume Interval: sample every xxx,xx m
3
by flow
signal (pulse or current).
Time Interval: sample every xx minutes.
Batch: sample every xx minutes when the pulse
input is activated.
Sample interval
Enter the value of the desired interval.
Error sample max
Enter the max. number of consecutive error
samples to trip the sample alarm.
Sample volume
Enter the sample volume.
Purge-, Max suction-,
Dose- (VS only) &
Activation Time (ILS
only)
Purge time:
time to clean the suction tube.
Max suction time
: maximum time to reach the
level pins. Dose time: time to dose the sample.
Activation time: time to activate the ILS sampler.
3.5.4
Changing Distributor Settings
Select submenu
Distributor Settings
and press
Enter
.
Select from the following parameters. Press
Enter
to change its value.
Parameter
Description
Turn Time
Enter the clock time the distributor needs to
change the sample outlet to the next container.
Turn Daily
Select the days the distributor needs to turn,
Everyday, Skip Sunday, or Off(don’t turn on the
Turn time)
Turn Interval
Enter the desired time interval for the distributor
to turn (0=off).
Container config.
Enter the number of containers & container
volume.
Overflow protect.
Select what action to take when the container
starts to overflow: Off (continue program), Next
container, Stop sampling (& restart when the
turn time is reached).
Summary of Contents for omy
Page 44: ...Installation User Manual Sampling Systems Version 140618 UK Page 44 ...
Page 45: ...Installation User Manual Sampling Systems Version 140618 UK Page 45 ...
Page 46: ...Installation User Manual Sampling Systems Version 140618 UK Page 46 ...
Page 47: ...Installation User Manual Sampling Systems Version 140618 UK Page 47 ...
Page 48: ...Installation User Manual Sampling Systems Version 140618 UK Page 48 ...