PUMP TEMP: **
℃
RATING: **000rpm
LEVITATION
Change speed by turning
VR1 volume
The massage displayed can
be changed by pressing the
“SELECT” switch.
Determine speed by
pressing “START” switch
< LCD Message >
<Operation>
SELECT Switch
SELECT Switch
SELECT Switch
Pump’s state
*1
Pump’s internal temperature
Set value of rated speed
Rated speed setting mode
SELECT Switch
SET = ***00rpm
DSW -B No.1 ON
◇
The rated speed returns to 48,000 rpm (initial setting) by setting dip sw itch 1 of
DSW-B to "OFF."
Figure 9.2 LCD Message in Rated Speed Setting Mode
*1: The STP pump’s state displayed depends on the pump’s operation state.
Examples) “LEVITATION” (LEVITATION state)
“ACCEL: ***00prm” (ACCELERATION state)
“NORMAL: ***00rpm” (NORMAL OPERATION state)
“BRAKE: ***00rpm” (BRAKE state)