11.3.2 Replacing the Fuses
◇
Remove the probable causes of the blow n fuse (Ex. a problem w ith the remote
input signal, and other) before replacing the fuse.
◇
Alw ays use specified and rated fuses.
◇
When you use parts other than those recommended in the above table, contact
Service Office.
◇
If fuses are blow n frequently even after they are replaced, contact Service
office.
◇
Tw o fuses are attached for replacement.
The following are protective fuses used in the STP control.
F1, F2: 250 V, 10 A
For I/O TB1 remote input/output terminal protection.
For power input (1) and (2) and power output (3) and (4) protection.
Table 11.5 Recommended Spare Parts
Product name
Type
Specifications
Fuses for F1, F2
MF60-NR10 10A, 250 V
(Mfr.: TOYO FUSE CO., LTD)
10A, 250 V, Normal acting type
φ
6.35
×
30 mm