8. Fault finding
Fault
The controller LEDs do not flash for 0.5 seconds when the system is switched on
Cause
The electrical supply has failed.
Remedy
Make sure that the electrical supply is set to on and the fuses (and current limiting
devices) are not tripped.
Fault
The pump does not start after a parallel start command is supplied
Cause
Check that the electricity supply is on. Check if the fail output is active.
Remedy
If there is a fail signal, check if the red alarm LED is flashing. If yes, refer to
Flashing error codes on page 77
. If the power is supplied and fail signal is not given, but
the rotor does not rotate, the fault is in the pump.
Fault
The pump does not start after a serial start command is sent
Cause
Check whether the pump returns a reply to the start command.
Remedy
If there is no reply then perform the checks given under
otherwise perform the checks under
The pump does not start after a manual start is
.
Fault
The pump does not start after a manual start is sent
Cause
Check that the electricity supply is on and whether the fail output is active.
Remedy
If there is a fail signal, check whether the red alarm LED is flashing. If it is, refer to
Table: Flashing error codes on page 77
. If power is supplied, there is no fail signal and
the pump still does not start then there is a fault with the pump.
Fault
The pump does not start after power is applied in auto run mode
Cause
Check that the electricity supply is on, that auto run mode is enabled and whether
the fail output is active.
Remedy
If there is a fail signal, check whether the red alarm LED is flashing. If it is, refer to
Table: Flashing error codes on page 77
. If power is supplied, there is no fail signal, the
auto run LED in on and the pump still does not start then there is a fault with the
pump.
Fault
The pump does not respond in multi-drop mode
Cause
The multi-drop is disabled.
Remedy
Make sure the pump has a multi-drop address and that commands are sent using the
multi-drop protocol.
Fault
No serial comms
Cause
No electrical supply, loss of serial link or serial enable.
Page 72
A65901880_D - Fault finding
Summary of Contents for nRVi
Page 1: ...Rotary Vane Pumps nRVi INSTRUCTION MANUAL A65901880_D Original instructions edwardsvacuum com...
Page 6: ...This page has been intentionally left blank...
Page 97: ...Page 93 A65901880_D Service...
Page 98: ...This page has been intentionally left blank...
Page 99: ...This page has been intentionally left blank...
Page 100: ...edwardsvacuum com...