Zero Master Calibration
1)
From RUN mode, press the CAL button.
2)
Rotate the ENTER button until "ZERO" is displayed and then press ENTER. "INS
ZERO" will be displayed.
NOTE: All checks that are programmed with a ZERO master will be turned "ON"
by default and will therefore be gaged at the same time. If the ZERO masters are to
be gaged separately then the checks must be turned "OFF" manually from the "CHK
SEL" menu.
3)
Place the ZERO master into gage position and press ENTER. "GAGE ZERO" will be
displayed.
4)
Press the ENTER button. The ZERO master will be gaged and then the display will
update to "MASTERED" if the calibration was successful. If the calibration was
successful the gage is ready to measure parts, see
Operation
. Remove the ZERO
master and press RUN . If the calibration was not completed successfully then see the
Troubleshooting section of this manual.
Summary of Contents for Accu-Setter II
Page 2: ...A Accu Setter II User s Manual Edmunds GAGES...
Page 16: ...2 3 2 3 Overall E9300Unit Figure 2 1 Accusetter II Basic Unit E9300...
Page 18: ...2 5 Figure 2 2 Front Panel...
Page 20: ...2 7 Figure 2 3 Rear Panel...
Page 32: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Page 39: ...3 11 An example of various multipliers as related to diameter measurement follows...
Page 40: ...3 12 The last consideration to complete the formula is the gaging function See examples below...
Page 43: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 15...
Page 44: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 16...
Page 50: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 22...
Page 51: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 23...
Page 59: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 31...
Page 60: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 32...
Page 90: ...4 5 Figure 4 3 A E Block...