4-26
Auto Air Setup
NOTE: The air check(s) mag and zero must be properly setup and the gage
calibrated before setting up the auto air shutoff.
1) From Run mode press the ENTER button and then rotate the ENTER button until
"SYSTEM" is displayed and then press ENTER.
2) Rotate the ENTER button until "AUTO AIR" is displayed and then press ENTER.
3) Rotate the ENTER button until "5", "10", or "15" seconds is displayed and press
ENTER to select the time delay before auto air shutoff.
4) Press "RUN" to return to run mode.
5) After the programmed delay the air will shutoff and the display on the column will
flash "WAIT" for several seconds and then display "AIR OFF".
NOTE: During the
time that the display is flashing "WAIT" DO NOT reinsert the air tool into the
part.
6) Close the flow control on the air shutoff valve so that a minimal amount of air is
allowed to flow to the air tooling.
7) Place the air tooling in the least material condition master. For an air plug the LMC
master is the max master and for an air ring the LMC master is the min master.
8) If the column readout detects two light bars of deviation when the air tool is loaded in
the master it will automatically turn on the full air flow to the gage and a
measurement can be made. If the column readout does not detect two light bars of
deviation when the air tool is loaded in the master then slowly open the flow control
valve on the air shutoff valve until a two bar deviation is detected and the air turns on.
9) Remove the air tool from the LMC master. After the programmed delay the air will
shutoff and the display on the column will flash "WAIT" for several seconds and then
display "AIR OFF".
NOTE: During the time that the display is flashing
"WAIT" DO NOT reinsert the air tool into the part.
10) Load the air tool back in the LMC to ensure the air turns on. If it does not turn on
open the flow control until the air tur ns on.
11) Repeat steps 9 and 10 until the air turns on and off correctly.
NOTE:
If the gage is
used to measure parts with greater clearance than the LMC master it may be
necessary to open the flow control valve more to ensure that the column detects a two
light bar deviation when the air tool is loaded.
Summary of Contents for Accu-Setter II
Page 2: ...A Accu Setter II User s Manual Edmunds GAGES...
Page 16: ...2 3 2 3 Overall E9300Unit Figure 2 1 Accusetter II Basic Unit E9300...
Page 18: ...2 5 Figure 2 2 Front Panel...
Page 20: ...2 7 Figure 2 3 Rear Panel...
Page 32: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Page 39: ...3 11 An example of various multipliers as related to diameter measurement follows...
Page 40: ...3 12 The last consideration to complete the formula is the gaging function See examples below...
Page 43: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 15...
Page 44: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 16...
Page 50: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 22...
Page 51: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 23...
Page 59: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 31...
Page 60: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 32...
Page 90: ...4 5 Figure 4 3 A E Block...