PROGRAMMING - MULTIPLE FIXTURE MEASUREMENT
3-37
Class (Optional)
1)
Rotate the ENTER button to display "CLASS" and press the ENTER button.
"EQUAL" or "SELECT" will display.
2)
Rotate the ENTER button to display your choice and press the ENTER button to select.
"QTY" with a numerical value will display.
3)
Rotate the ENTER button left or right until desired number of classes is displayed. (39
maximum).
4)
Press the ENTER button. "STRT PT" will display.
5)
Rotate the ENTER button left or right to set the starting point of the first class.
6)
Press the ENTER button. "CLASS" with a numerical value will display.
7)
Rotate the ENTER button to set the ending point of the first class.
8)
Press the ENTER button to select. "CLASS 02" will display if more than one class has
been selected. Note: If you chose EQUAL, the Accu-Setter will automatically program
the remaining classes. If you chose SELECT, rotate the ENTER button to set the
ending point of the next class. Press the enter button to select. The next class number
will display. Repeat procedure until all classes are programmed.
9)
Either "SELECT" or "EQUAL" will display.
10)
Press the RST button twice until "LIMITS" is displayed.
Master
This menu allows the operator to select which masters will be used to calibrate the gage
and to program the correct master values. For example a two master system uses a MAX
and MIN master while a single master system would use only the ZERO master. The
auxiliary (AUX) master is used to calibrate a TIR check mastered with a single master.
Refer to the Accusetter setup guide to determine what master or masters will be used to
calibrate the gage and refer to the master calibration reports for the actual master values.
1)
Rotate the ENTER button until "MASTER" is displayed and then press the ENTER
button.
2)
Rotate the ENTER button until "MAX", "MIN", "AUX", or "ZERO" is displayed and
then press enter.
3)
Rotate the ENTER button to select "ON" or "OFF" and then press ENTER.
4)
NOTE: If the auxiliary master ("AUX") is turned "ON" then "MAX" and "MIN" are
automatically turned "OFF" and if "MAX" or "MIN" is turned "ON" then "AUX" is
automatically turned "OFF".
5)
If "ON" was selected then rotate the ENTER button to drive the bargraph to the master
value listed on the master certification report and then press ENTER and return to step
2. If "OFF" was selected then return to step 2.
6)
After all required masters are programmed press RST to return to the define menu.
Summary of Contents for Accu-Setter II
Page 2: ...A Accu Setter II User s Manual Edmunds GAGES...
Page 16: ...2 3 2 3 Overall E9300Unit Figure 2 1 Accusetter II Basic Unit E9300...
Page 18: ...2 5 Figure 2 2 Front Panel...
Page 20: ...2 7 Figure 2 3 Rear Panel...
Page 32: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Page 39: ...3 11 An example of various multipliers as related to diameter measurement follows...
Page 40: ...3 12 The last consideration to complete the formula is the gaging function See examples below...
Page 43: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 15...
Page 44: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 16...
Page 50: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 22...
Page 51: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 23...
Page 59: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 31...
Page 60: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 32...
Page 90: ...4 5 Figure 4 3 A E Block...