417102012_LMI02.doc
- 85 -
Rev. 9-07.08
8
Checking the conductivity display
The conductivity display of the LMI02 can be checked with the conductivity simulator for
LMIT08 (item no. 289190) or with a calibration box for multronic (item no. 255195). It is
also possible to check this by dipping the conductivity probe into a sample solution of
known conductivity.
CAUTION
The conductivity probe should be brought to 20°C in order to check the conductivity
display and the current output.
When using a simulator, the
↵
key can be pressed to show the uncompensated
conductivity value (a * symbol appears in front of the conductivity display). The current
output is also adjusted to the uncompensated value.
Current output on LMI02
LMI02
measuring
range
With simulator for LMIT08
item no. 289190
Resistance value (label)
Conductivity
display
on LMI02
0..20 mA
4..20 mA
0... 20 mS/cm)
383
Ω
(20.0 mS)
18.0 mS/cm)
18.0 mA
18.4 mA
0…200 mS/cm)
38.3
Ω
(200 mS)
180 mS/cm)
18.0 mA
18.4 mA
Current output on LMI02
LMI02
measuring
range
With simulator for Multronic
item no. 255195
Resistance value (label)
Conductivity
display
on LMI02
0..20 mA
4..20 mA
0... 20 mS/cm)
345
Ω
(20.0 mS)
20.0 mS/cm)
20.0 mA
20 mA
0…200 mS/cm)
34.5
Ω
(200 mS)
200 mS/cm)
20.0 mA
20 mA
The adjustment of the conductivity display is described in chapter
of the LMI02
User’s Manual.
fig. 8.1
Loop resistance
(simulations resistance)
Summary of Contents for LMI 02
Page 47: ...417102012_LMI02 doc 47 Rev 9 07 08 13 Anhang 13 1 Anwendungsbeispiele Abb 13 1 Abb 13 2...
Page 51: ...417102012_LMI02 doc 51 Rev 9 07 08 13 3 Programmier bersicht 13 3 1 AS Mode...
Page 52: ...417102012_LMI02 doc 52 Rev 9 07 08 13 3 2 CIP Mode...
Page 53: ...417102012_LMI02 doc 53 Rev 9 07 08 14 Konformit tserkl rung...
Page 93: ...417102012_LMI02 doc 93 Rev 9 07 08 13 Appendix 13 1 Application examples fig 13 1 fig 13 2...
Page 97: ...417102012_LMI02 doc 97 Rev 9 07 08 13 3 Programming overview 13 3 1 AS Mode...
Page 98: ...417102012_LMI02 doc 98 Rev 9 07 08 13 3 2 CIP Mode...
Page 99: ...417102012_LMI02 doc 99 Rev 9 07 08 14 Declaration of Conformity...
Page 139: ...417102012_LMI02 doc 139 Rev 9 07 08 13 Annexe 13 1 Exemples d application Fig 13 1 Fig 13 2...
Page 143: ...417102012_LMI02 doc 143 Rev 9 07 08 13 3 Tableau de programmation 13 3 1 AS Mode...
Page 144: ...417102012_LMI02 doc 144 Rev 9 07 08 13 3 2 CIP Mode...
Page 145: ...417102012_LMI02 doc 145 Rev 9 07 08 14 D claration de conformit...