SPECIFICATION
Two instrument range options are available. See below.
2.1 Ranges
Ranges
Gas mix
Display resolution
Option 1
Range 1
0…5% max.
H
2
in Cl
2
and background gas
0.01%
Range 2
0…100%
Cl
2
in H
2
and background gas
0.1%
Option 2
Range 1
0…10% max.
H
2
in Cl
2
and background gas
0.01%
Range 2
0…100%
Cl
2
in H
2
and background gas
0.1%
2.2 Stability
Better than 1% f.s.d. over the operating temperature range
Better than 1% f.s.d./month
2.3 Accuracy
±1% f.s.d. or ±2% f.s.d. depending on calibration method
2.4 Repeatability
Better than 1% f.s.d.
2.5 Sample flow
Between 100 to 400 ml/min
2.6 Sample temperature range
0°C to +60°C (non-condensing)
2.7 Sample pressure
Atmospheric (nominally) - set by vent pressure
Maximum pressure - 1 barg
2.8 Ambient operating temperature range
Katharometer/Reactor:
+5°C to +55°C
Electronics:
+5°C to +45°C (0-90% R.H. non-condensing)
2.9 Storage temperature range
0°C to 55°C (0-90% R.H. non-condensing)
2.10 Displays
Instrument: 4 line L.C.D.
Reactor controller: 4 digit L.E.D.