background image

10

INM MTL 130-0202 Rev 4

OPERATION

4.1  First checks

It is advisable to make a thorough check of all connections – particularly the pipe work – 
before applying any sample.

WARNING

Chlorine is a highly toxic and corrosive gas and it is of utmost importance that 
there are no leaks. In particular, ensure that all unions are fully tightened and 
that the fragile reactor tube is not damaged.

4.2  Flowing the sample

With these checks complete, power may be applied to the instrument.  Following the 
initialisation screen, the display will indicate that it is waiting for the reactor to reach 
temperature. When the correct operating temperature is reached, the analyser will enter 
normal measurement mode automatically.  The sample may now be applied.  A sample 
flow of between 100 and 300 ml/min is suggested, although flows up to 400ml/min may be 
applied.  (100ml/min = 0.212cuft/hr)

Once the sample pipelines have been purged the instrument display will show the hydrogen 
and chlorine concentrations.  A warm-up time of approximately 15 minutes after the reactor 
has reached its operating temperature should be allowed in order to obtain the best accuracy. 
The instrument is supplied fully tested and calibrated and may be used immediately.

4.3  Configuring the analyser 

This enables the user to step through a range of display screens and define (program) specific 
parameters, such as alarm levels, ranges, etc. The screens also provide data on the model 
and the software revision of the analyser. The following table describes the function and 
parameters available at each screen. There is also a diagram of the screen levels followed by 
a description of how to set the parameters.

4.3.1  Parameters 

Screen 

Function

Parameter

Description

1

Alarm 1 (Hydrogen 

Concentration)

Alarm Level

Adjustable from 0 to 5% or 0 to 10% (model dependant) in 0.01% steps

Hysteresis

Adjustable from 0 to 2% of span in 0.01% steps

Alarm Mode

Can be set to High, Low or Off

2

Alarm 2 (Chlorine 

Concentration)

Alarm Level

Adjustable from 0 to 100% in 0.1% steps

Hysteresis

Adjustable from 0 to 20% of span in 0.01% steps

Alarm Mode

Can be set to High, Low or Off

3

Hydrogen Span 

Calibration Level

Level

Concentration level adjustable from 1.50 to 2.00% in 0.01% steps

4

Chlorine Span 

Calibration Level

H

Cl

2

NOTE: The facility to set the hydrogen concentration in this screen is to facilitate 
calibration on a plant sample when the hydrogen concentration is known from other 
analysers.  If pure chlorine OR pure argon (the surrogate for chlorine) is used for chlorine 
span calibration, the hydrogen level in this screen MUST BE SET TO ZERO.

5

4/20mA function

Cal. 
Fault

Define the action of the analogue output, when the analyser is in calibration or fault 
conditions. Set either to Follow (track) the display or Hold the last value immediately 
prior to the condition occurring.

6

H

2

  4/20mA

20mA

Set top H

2

 scale value. Adjustable from 1.00% to 5.00% in 0.01% steps, or 2.00% to 

10.00% in 0.01% steps. (model dependant)

7

Filter

Level

A digital filter to ‘smooth’ rapid display changes. Set level between 0 and 9 (represents 
factor of 0.0 to 0.9).

Displayed change = Actual change x (1 – factor)

e.g. Change of 50% up to 70% with Level = 5 would result in 20% x (1 – 0.5) = 10%; 
hence next display is 50 + 10 = 60%

8

Samples, ranges 

and output

Displays analyser type, ranges measurable, and type of output

9

Sensor & software

Displays sensor serial number, software and version data.

Summary of Contents for KK650

Page 1: ...KK650 MTL hydrogen and chlorine gas analyser January 2017 INM MTL 130 0202 Rev 4 Instruction manual MTL gas analysers systems...

Page 2: ...LARATION OF CONFORMITY A printed version of the Declaration of Conformity has been provided separately within the original shipment of goods However you can find a copy of the latest version at http w...

Page 3: ...unit 4 2 17 Katharometer reactor unit 4 3 INSTALLATION 5 3 1 Unpacking and visual checking 5 3 2 Mounting 5 3 3 Electrical connections 5 3 4 Installation requirements for EMC 5 3 5 Sample conditioning...

Page 4: ...iv INM MTL 130 0202 Rev 4 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 5: ...n 1 3 How it works First the thermal conductivity of the plant sample as delivered is measured Next the sample is passed through a heated reactor tube where the hydrogen is reacted with the excess of...

Page 6: ...used to give general information to ensure correct operation 1 5 Information Waste Electrical and Electronic Equipment directive WEEE 2002 96 EC RoHS directive 2002 95 EC WARNING This equipment must...

Page 7: ...erating temperature range Better than 1 f s d month 2 3 Accuracy 1 f s d or 2 f s d depending on calibration method 2 4 Repeatability Better than 1 f s d 2 5 Sample flow Between 100 to 400 ml min 2 6...

Page 8: ...n is NC see Figure 6 2 13 Concentration alarms Two user configurable Alarm outputs are provided 1 Programmable alarm for Cl2 2 Programmable alarm for H2 Both are volt free SPDT relay contacts rated at...

Page 9: ...he relevant parts of IEC 60947 1 and IEC 60947 3 It must be marked to indicate this function and show ON and OFF positions Wiring should conform to local codes Only the live conductor has an internal...

Page 10: ...mped System designers and installers should consult Perma Pure for detailed advice before installing any system www permapure com a Sample conditioning Aspirated version b Sample conditioning Pumped S...

Page 11: ...thin 0 1 barg for measurement accuracy Figure 3 Reactor Tube FRAGILE Four mounting holes on 220mm x 130mm centres suitable for M6 screws WARNING These parts run HOT in normal operation Outline of heat...

Page 12: ...the incoming AC power are found in the lowest part of the enclosure as can be seen in Figure 5 below Electronics Control unit mounted on enclosure backplate Reactor Ready Reactor fuse T 5 A View Edit...

Page 13: ...minated Its connecting cable should therefore be routed to assist in its removal Allow sufficient space around the katharometer reactor assembly for withdrawal of the quartz reactor tube to the right...

Page 14: ...the screen levels followed by a description of how to set the parameters 4 3 1 Parameters Screen Function Parameter Description 1 Alarm 1 Hydrogen Concentration Alarm Level Adjustable from 0 to 5 or 0...

Page 15: ...TS Luton England Hydrogen in Chlorine Analyser 1 Press the VIEW key to access screen 1 Press the up arrow key to scroll in this direction Press the down arrow key to scroll in this direction View Mode...

Page 16: ...turn to view mode displaying the new parameters and un freezing the outputs If no keys are pressed for any 30 second period while in Edit mode an automatic time out takes the instrument back into meas...

Page 17: ...was 96 chlorine and 0 3 hydrogen this is what would need to be entered in Configuration Screen 4 NOTE The hydrogen concentration set in Configuration Screen 4 is to enable the analyser to compensate...

Page 18: ...displayed will be the argon equivalent for that particular katharometer This figure varies from 33 to 35 The precise point varies slightly from one katharometer to another and the exact figure is a pa...

Page 19: ...IT for Zero 2 H span 3 Cl span VIEW to exit 2 Calibration mode Surrogate Non surrogate VIEW to exit Pass Dry air Cl MEAS When ready VIEW to exit 2 Screen colours Surrogate gas calibration Non surrogat...

Page 20: ...on Incorrect calibration The first two are normally picked up by the display of a fault code the last is cured by a recalibration 6 2 Fault codes Fault code Meaning 1 Vo1 has measured 4093 A D counts...

Page 21: ...ge signal that indicates the katharometer s temperature Vo1 is the internal voltage signal that is a function of the difference between the two katharometer sense signals at terminals 1 2 and 3 4 in F...

Page 22: ...fuse 2 Replace as necessary Fault code between 1 and 32 is displayed see manual sec 5 2 1 Unexpected gas applied e g high hydrogen content more than 60 over instrument range 1 Pass air through the ana...

Page 23: ...Katharometer Identity Module KIM 6 4 Checking the katharometer The simplest way to identify a possible katharometer fault is to measure the volts appearing across the measurement elements 6 4 1 Checki...

Page 24: ...blue plastic PFA coupling nuts of the sample inlet and outlet and pull the connecting pipes to one side ENSURE THAT THE FRONT FERRULES see fig 3 ARE EITHER REMOVED AND PUT IN A SAFE PLACE OR RETAINED...

Page 25: ...wn so that the water irrigates the sensor element Five minutes of flowing tap water should suffice 3 After the tap water wash shake out as much water as possible and rinse several times with distilled...

Page 26: ...ck the resistance across the ends of the cable This should be in the order of 4 ohms If it is substantially higher then the reactor assembly will need replacing If the voltage is at or near zero check...

Page 27: ...o a suitable torque value 3 Refit the earth wire with the others on the backplate grounding stud and replace the securing washer and nut Tighten to a suitable torque value The instrument will now need...

Page 28: ...trozavodskaya Str 33 Building 4 Moscow 107076 Russia Tel 7 495 981 3770 Fax 7 495 981 3771 E mail mtlrussia eaton com SINGAPORE Cooper Crouse Hinds Pte Ltd No 2 Serangoon North Avenue 5 06 01 Fu Yu Bu...

Reviews: