background image

INM MTL 130-0202 Rev 4

23

6.9.2  Removal and replacement

The power supply to the instrument must be completely disconnected.

The stainless steel electronics unit mounts to the backplate of the enclosure on four captive 
studs, two each side, and is secured to each with a nut, a locking washer and a plain washer - 
see Figure 4 (page 8) for details.

1.  Remove the nut on the backplate grounding stud so that the earth wire connected to   
 

the steel casing of the control unit can be detached easily at the end of the next step.

2.  Support the control unit while removing the nuts and washers from the studs. The  
 

electronics unit can then be detached from the connectors at the rear together with its  

 

earth wire.

Replacement is the reversal of removal. Before fitting the replacement electronics ensure that 
it is configured for the same supply voltage as the one it is replacing, this can be determined 
from the label on the side of the electronics unit.  If in doubt, contact your local MTL Gas 
sales office who will require the serial numbers of both the old and the new units to check 
the details.

1.  Thread the earth wire behind the lower cover then carefully align the new module into  
 

position on the mounting studs. Carefully press the module into position to fully engage the  

 

rear connectors.

2.  While supporting the unit in place, fit a plain washer and a locking washer over one of the  
 

top studs and spin a nut onto it until finger-tight. Do the same on the other top stud to  

 

secure the unit. Fit the lower washers and nuts in a similar manner then tighten all four  

 

nuts to a suitable torque value.

3.  Refit the earth wire, with the others, on the backplate grounding stud and replace the  
 

securing washer and nut. Tighten  to a suitable torque value.

The instrument will now need to be calibrated and any user configuration settings will need to 
be re-entered.  See sections 5.3 and 6 for details.

6.10  Restoring factory defaults

This procedure may be used where mis-calibration is producing unacceptable errors and re-
calibration is not possible – e.g. no gas available.

This procedure should be carried out by qualified service personnel only, and only as indicated 
by the fault finding requirements set out in the trouble-shooting chart in section 7.3.

6.10.1   Tools required

None

6.10.2  Instructions

Disconnect the instrument from its power supply and leave for 30 seconds.  Whilst keeping 
the 

Cal

 button depressed, apply power to the instrument.  Continue to hold the 

Cal

 button 

until a ‘Calibration data lost’ message appears on the display.

Connect a dry air source to the instrument and flow air at approx 200 ml /min through the 
sensor and check the chlorine and hydrogen zero accuracy against the table below.  If the 
readings are outside the limits shown, it is likely that the katharometer sensor is contaminated 
and will need cleaning.  See section 7.6 for the cleaning procedure. Should the cleaning not 
restore operation to within limits shown below, the sensor and its KIM should be replaced.

If the results are within the limits shown here, recalibrate the analyser and restore the 
customer settings from the noted values above.

6.10.3  Acceptable limits 

Reading

Error - %Absolute

Chlorine

± 8%

Hydrogen

± 1%

These limits refer to the zero reading accuracy at room temperature after restoring factory 
settings but before performing a gas calibration.

Summary of Contents for KK650

Page 1: ...KK650 MTL hydrogen and chlorine gas analyser January 2017 INM MTL 130 0202 Rev 4 Instruction manual MTL gas analysers systems...

Page 2: ...LARATION OF CONFORMITY A printed version of the Declaration of Conformity has been provided separately within the original shipment of goods However you can find a copy of the latest version at http w...

Page 3: ...unit 4 2 17 Katharometer reactor unit 4 3 INSTALLATION 5 3 1 Unpacking and visual checking 5 3 2 Mounting 5 3 3 Electrical connections 5 3 4 Installation requirements for EMC 5 3 5 Sample conditioning...

Page 4: ...iv INM MTL 130 0202 Rev 4 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 5: ...n 1 3 How it works First the thermal conductivity of the plant sample as delivered is measured Next the sample is passed through a heated reactor tube where the hydrogen is reacted with the excess of...

Page 6: ...used to give general information to ensure correct operation 1 5 Information Waste Electrical and Electronic Equipment directive WEEE 2002 96 EC RoHS directive 2002 95 EC WARNING This equipment must...

Page 7: ...erating temperature range Better than 1 f s d month 2 3 Accuracy 1 f s d or 2 f s d depending on calibration method 2 4 Repeatability Better than 1 f s d 2 5 Sample flow Between 100 to 400 ml min 2 6...

Page 8: ...n is NC see Figure 6 2 13 Concentration alarms Two user configurable Alarm outputs are provided 1 Programmable alarm for Cl2 2 Programmable alarm for H2 Both are volt free SPDT relay contacts rated at...

Page 9: ...he relevant parts of IEC 60947 1 and IEC 60947 3 It must be marked to indicate this function and show ON and OFF positions Wiring should conform to local codes Only the live conductor has an internal...

Page 10: ...mped System designers and installers should consult Perma Pure for detailed advice before installing any system www permapure com a Sample conditioning Aspirated version b Sample conditioning Pumped S...

Page 11: ...thin 0 1 barg for measurement accuracy Figure 3 Reactor Tube FRAGILE Four mounting holes on 220mm x 130mm centres suitable for M6 screws WARNING These parts run HOT in normal operation Outline of heat...

Page 12: ...the incoming AC power are found in the lowest part of the enclosure as can be seen in Figure 5 below Electronics Control unit mounted on enclosure backplate Reactor Ready Reactor fuse T 5 A View Edit...

Page 13: ...minated Its connecting cable should therefore be routed to assist in its removal Allow sufficient space around the katharometer reactor assembly for withdrawal of the quartz reactor tube to the right...

Page 14: ...the screen levels followed by a description of how to set the parameters 4 3 1 Parameters Screen Function Parameter Description 1 Alarm 1 Hydrogen Concentration Alarm Level Adjustable from 0 to 5 or 0...

Page 15: ...TS Luton England Hydrogen in Chlorine Analyser 1 Press the VIEW key to access screen 1 Press the up arrow key to scroll in this direction Press the down arrow key to scroll in this direction View Mode...

Page 16: ...turn to view mode displaying the new parameters and un freezing the outputs If no keys are pressed for any 30 second period while in Edit mode an automatic time out takes the instrument back into meas...

Page 17: ...was 96 chlorine and 0 3 hydrogen this is what would need to be entered in Configuration Screen 4 NOTE The hydrogen concentration set in Configuration Screen 4 is to enable the analyser to compensate...

Page 18: ...displayed will be the argon equivalent for that particular katharometer This figure varies from 33 to 35 The precise point varies slightly from one katharometer to another and the exact figure is a pa...

Page 19: ...IT for Zero 2 H span 3 Cl span VIEW to exit 2 Calibration mode Surrogate Non surrogate VIEW to exit Pass Dry air Cl MEAS When ready VIEW to exit 2 Screen colours Surrogate gas calibration Non surrogat...

Page 20: ...on Incorrect calibration The first two are normally picked up by the display of a fault code the last is cured by a recalibration 6 2 Fault codes Fault code Meaning 1 Vo1 has measured 4093 A D counts...

Page 21: ...ge signal that indicates the katharometer s temperature Vo1 is the internal voltage signal that is a function of the difference between the two katharometer sense signals at terminals 1 2 and 3 4 in F...

Page 22: ...fuse 2 Replace as necessary Fault code between 1 and 32 is displayed see manual sec 5 2 1 Unexpected gas applied e g high hydrogen content more than 60 over instrument range 1 Pass air through the ana...

Page 23: ...Katharometer Identity Module KIM 6 4 Checking the katharometer The simplest way to identify a possible katharometer fault is to measure the volts appearing across the measurement elements 6 4 1 Checki...

Page 24: ...blue plastic PFA coupling nuts of the sample inlet and outlet and pull the connecting pipes to one side ENSURE THAT THE FRONT FERRULES see fig 3 ARE EITHER REMOVED AND PUT IN A SAFE PLACE OR RETAINED...

Page 25: ...wn so that the water irrigates the sensor element Five minutes of flowing tap water should suffice 3 After the tap water wash shake out as much water as possible and rinse several times with distilled...

Page 26: ...ck the resistance across the ends of the cable This should be in the order of 4 ohms If it is substantially higher then the reactor assembly will need replacing If the voltage is at or near zero check...

Page 27: ...o a suitable torque value 3 Refit the earth wire with the others on the backplate grounding stud and replace the securing washer and nut Tighten to a suitable torque value The instrument will now need...

Page 28: ...trozavodskaya Str 33 Building 4 Moscow 107076 Russia Tel 7 495 981 3770 Fax 7 495 981 3771 E mail mtlrussia eaton com SINGAPORE Cooper Crouse Hinds Pte Ltd No 2 Serangoon North Avenue 5 06 01 Fu Yu Bu...

Reviews: