background image

INM MTL 130-0202 Rev 4

13

CALIBRATION 

5.1  General notes

Provided the katharometer does not become contaminated, instruments can be expected 
to hold their calibration repeatability to better than 0.02% hydrogen for a lengthy period. 
Chlorine accuracy varies according to air and hydrogen concentrations.  In the 90 to 100% 
range it is approximately 0.5% going out to about 1.5% at lower concentrations with high 
hydrogen.   If the ratio of oxygen to nitrogen varies then a small error is introduced into the 
chlorine measurement.  For every 1% oxygen variation the apparent chlorine concentration is 
changed by 0.07% (higher oxygen gives higher chlorine).

5.2  Gases required

To avoid the need to handle hazardous gases, calibration may be performed using benign 
surrogate gases.  The surrogate gases used are; 1.5 - 2% hydrogen in air (50% LEL mixture) 
as a substitute for hydrogen in chlorine, and 100% argon in place of chlorine.

Another, even more convenient, way to calibrate the chlorine span of the analyser is to 
use the process sample gas - assuming its composition is reliably known from other forms 
of analysis. Configuration Screen 4 (see configuration diagram in section 3), allows both 
the chlorine and hydrogen concentrations (see following Note) to be set for the gas used 
for calibration. For example, if the process sample analysis was 96% chlorine and 0.3% 
hydrogen, this is what would need to be entered in Configuration Screen 4.

NOTE

The hydrogen concentration set in Configuration Screen 4 is to enable the analyser to 

compensate for its presence when calibrating the chlorine span on process sample – it is 

NOT a part of calibrating the hydrogen span.  The latter is performed using hydrogen in 

air and specifying the hydrogen concentration in Configuration Screen 3.  The hydrogen 

concentration set in Screen 4 must be zero if pure gases (i.e. 100% chlorine or 100% 

argon) are used to calibrate the chlorine span.

The calibration process must be performed in the order of

 

1.  Zero calibration (air)

 

2.  Hydrogen span (2% hydrogen in air)

 

3.  Chlorine span (100% chlorine, or 100% argon or process gas if analysis is known)

Air is the ‘zero’ gas, and is common to both surrogate and non-surrogate calibration 
processes.  It is possible to calibrate using both surrogate and non-surrogate gases.  So it is 
possible to zero on air, set the hydrogen span using hydrogen in air, and then set the chlorine 
span using either pure chlorine or argon, or the process gas if its analysis is reliably known 
and the hydrogen and chlorine concentrations are entered as described above.  For best 
accuracy, pure chlorine should be used for setting the chlorine span.

NOTE

Hydrogen in chlorine mixtures can only be generated on-the-spot because they are unstable.  

Outside of the laboratory the practical difficulties of making these mixtures rule them out of 

being used.

continued on next page

4.3.2  Edit mode

To edit a value, start by pressing the 

View

 button to enter the screen choice level. Use the 

arrow buttons to step through each screen and its options. At the chosen parameter, press 
the Edit button at which point the following screen is displayed.

This is to warn that all outputs will be stopped at their current value if the user goes further. 
Before continuing, the operator must be sure that no plant malfunction or safety problems 
will occur due to the ‘freezing’ of the analyser outputs.

When Edit mode is entered, a flashing cursor will appear on the first digit that may be altered. 
The keys increment or decrement the digit.  Once a digit is correct the next digit is selected 
by pressing the Edit key, and so on, until the final digit.  To skip a digit, press the Edit key 
twice.  When the Edit key is pressed on the final digit the display will show “Storing Data” 
momentarily and then return to view mode, displaying the new parameters and un-freezing 
the outputs.

If no keys are pressed for any 30 second period while in Edit mode an automatic time-out 
takes the instrument back into measurement mode and displays a screen to this effect. Any 
changes made during Edit mode prior to this are ignored and lost.

NOTE

When setting alarms the maximum hysteresis that can be set is 20% of the span of the 

particular channel.  It is important not to set the hysteresis to a level greater than the alarm 

point otherwise the alarm will never reset.

The following is a summary of the function of the keys when in View and Edit modes.

View mode (block cursor not displayed):-

 

return to the previous parameter or screen

advance to the next parameter or screen

View

exit View mode and return to Measurement mode

Edit

displays warning screen - press again for edit cursor.  See below

Edit mode (block cursor displayed):-

 

increment the digit under the cursor

decrement the digit under the cursor

View

exit Edit mode - return to View mode

Edit

advance to the next digit - or store entry (if the last digit has been changed)  
and return to View mode

Summary of Contents for KK650

Page 1: ...KK650 MTL hydrogen and chlorine gas analyser January 2017 INM MTL 130 0202 Rev 4 Instruction manual MTL gas analysers systems...

Page 2: ...LARATION OF CONFORMITY A printed version of the Declaration of Conformity has been provided separately within the original shipment of goods However you can find a copy of the latest version at http w...

Page 3: ...unit 4 2 17 Katharometer reactor unit 4 3 INSTALLATION 5 3 1 Unpacking and visual checking 5 3 2 Mounting 5 3 3 Electrical connections 5 3 4 Installation requirements for EMC 5 3 5 Sample conditioning...

Page 4: ...iv INM MTL 130 0202 Rev 4 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 5: ...n 1 3 How it works First the thermal conductivity of the plant sample as delivered is measured Next the sample is passed through a heated reactor tube where the hydrogen is reacted with the excess of...

Page 6: ...used to give general information to ensure correct operation 1 5 Information Waste Electrical and Electronic Equipment directive WEEE 2002 96 EC RoHS directive 2002 95 EC WARNING This equipment must...

Page 7: ...erating temperature range Better than 1 f s d month 2 3 Accuracy 1 f s d or 2 f s d depending on calibration method 2 4 Repeatability Better than 1 f s d 2 5 Sample flow Between 100 to 400 ml min 2 6...

Page 8: ...n is NC see Figure 6 2 13 Concentration alarms Two user configurable Alarm outputs are provided 1 Programmable alarm for Cl2 2 Programmable alarm for H2 Both are volt free SPDT relay contacts rated at...

Page 9: ...he relevant parts of IEC 60947 1 and IEC 60947 3 It must be marked to indicate this function and show ON and OFF positions Wiring should conform to local codes Only the live conductor has an internal...

Page 10: ...mped System designers and installers should consult Perma Pure for detailed advice before installing any system www permapure com a Sample conditioning Aspirated version b Sample conditioning Pumped S...

Page 11: ...thin 0 1 barg for measurement accuracy Figure 3 Reactor Tube FRAGILE Four mounting holes on 220mm x 130mm centres suitable for M6 screws WARNING These parts run HOT in normal operation Outline of heat...

Page 12: ...the incoming AC power are found in the lowest part of the enclosure as can be seen in Figure 5 below Electronics Control unit mounted on enclosure backplate Reactor Ready Reactor fuse T 5 A View Edit...

Page 13: ...minated Its connecting cable should therefore be routed to assist in its removal Allow sufficient space around the katharometer reactor assembly for withdrawal of the quartz reactor tube to the right...

Page 14: ...the screen levels followed by a description of how to set the parameters 4 3 1 Parameters Screen Function Parameter Description 1 Alarm 1 Hydrogen Concentration Alarm Level Adjustable from 0 to 5 or 0...

Page 15: ...TS Luton England Hydrogen in Chlorine Analyser 1 Press the VIEW key to access screen 1 Press the up arrow key to scroll in this direction Press the down arrow key to scroll in this direction View Mode...

Page 16: ...turn to view mode displaying the new parameters and un freezing the outputs If no keys are pressed for any 30 second period while in Edit mode an automatic time out takes the instrument back into meas...

Page 17: ...was 96 chlorine and 0 3 hydrogen this is what would need to be entered in Configuration Screen 4 NOTE The hydrogen concentration set in Configuration Screen 4 is to enable the analyser to compensate...

Page 18: ...displayed will be the argon equivalent for that particular katharometer This figure varies from 33 to 35 The precise point varies slightly from one katharometer to another and the exact figure is a pa...

Page 19: ...IT for Zero 2 H span 3 Cl span VIEW to exit 2 Calibration mode Surrogate Non surrogate VIEW to exit Pass Dry air Cl MEAS When ready VIEW to exit 2 Screen colours Surrogate gas calibration Non surrogat...

Page 20: ...on Incorrect calibration The first two are normally picked up by the display of a fault code the last is cured by a recalibration 6 2 Fault codes Fault code Meaning 1 Vo1 has measured 4093 A D counts...

Page 21: ...ge signal that indicates the katharometer s temperature Vo1 is the internal voltage signal that is a function of the difference between the two katharometer sense signals at terminals 1 2 and 3 4 in F...

Page 22: ...fuse 2 Replace as necessary Fault code between 1 and 32 is displayed see manual sec 5 2 1 Unexpected gas applied e g high hydrogen content more than 60 over instrument range 1 Pass air through the ana...

Page 23: ...Katharometer Identity Module KIM 6 4 Checking the katharometer The simplest way to identify a possible katharometer fault is to measure the volts appearing across the measurement elements 6 4 1 Checki...

Page 24: ...blue plastic PFA coupling nuts of the sample inlet and outlet and pull the connecting pipes to one side ENSURE THAT THE FRONT FERRULES see fig 3 ARE EITHER REMOVED AND PUT IN A SAFE PLACE OR RETAINED...

Page 25: ...wn so that the water irrigates the sensor element Five minutes of flowing tap water should suffice 3 After the tap water wash shake out as much water as possible and rinse several times with distilled...

Page 26: ...ck the resistance across the ends of the cable This should be in the order of 4 ohms If it is substantially higher then the reactor assembly will need replacing If the voltage is at or near zero check...

Page 27: ...o a suitable torque value 3 Refit the earth wire with the others on the backplate grounding stud and replace the securing washer and nut Tighten to a suitable torque value The instrument will now need...

Page 28: ...trozavodskaya Str 33 Building 4 Moscow 107076 Russia Tel 7 495 981 3770 Fax 7 495 981 3771 E mail mtlrussia eaton com SINGAPORE Cooper Crouse Hinds Pte Ltd No 2 Serangoon North Avenue 5 06 01 Fu Yu Bu...

Reviews: