NOTE
Care must be taken to ensure that the tubing etc, carrying the sample from the take-off
point to the katharometer entry is completely obscured from any light, otherwise the
analyser will report a lower hydrogen concentration for the sampled process gas. This
fact would of course be true for any other thermal conductivity analyser performing this
measurement.
Perma Pure dryers have been found to be very effective in drying the sample. Two optional
configurations of these are shown below - a) aspirated or b) pumped. System designers and
installers should consult Perma Pure for detailed advice before installing any system. (www.
permapure.com)
a) Sample conditioning - Aspirated version
b) Sample conditioning - Pumped System