Easytech s.r.l.
– Genu Iso/Plus User’s Manual
rev. C
Pag.
18
di
97
4.
CONNECTION TO GRID, FIRST TURN ON AND START
PROGRAM
4.1
CONNECTION TO GRID
Insert the power cord to device and to grid (100-240 V 50/60 Hz)
Make sure the Emergency Stop Button is in a raised position, if not, put it in its fully
raised position by turning it as shown by the arrows.
Turn on the device by using the green switch, it will light up.
CHECKS TO BE PERFORMED IF THE DEVICE WILL NOT POWER ON
CHECK
SOLUTION
Is the grid power p
resent in the device’
s
installation site?
- Contact your own electrician
Is the power cord correctly connected to
device? Is it intact?
- Connect the power cord whit care to
device. Check the power cord and
eventually replace it.
Is the main fuse good?
- Check the main fuse and replace if
necessary.
Summary of Contents for Isokinetic Genu Iso
Page 65: ...Easytech s r l Genu Iso Plus User s Manual rev C Pag 65 di 97 ...
Page 84: ...Easytech s r l Genu Iso Plus User s Manual rev C Pag 84 di 97 ...
Page 95: ......