Error Codes
5B6333 – 30SEP98
7
Non-Fatal Errors
Code
Description
Possible Malfunction
Action
E032
Developer tank fill error
This error will occur if the:
• DEVELOPER TANK:
– does not fill in 4 minutes in normal operation.
– does not fill in 15 minutes in the Tank Fill Mode.
– is empty and the operator does not select the Tank Fill Mode.
• REPLENISHMENT TANK is empty.
• REPLENISHMENT HOSE has an obstruction.
• TANKS of the PROCESSOR are filled with water during the initial
installation.
The following parts will be disabled:
• DEVELOPER REPLENISHMENT PUMP
• RECIRCULATION PUMP
• Temperature control for the fixer and developer
To prevent the error from occurring during the initial installation:
• add 240 mL (8 fl oz) of developer to the DEVELOPER TANK before you
fill the PROCESSOR with water.
• energize the RECIRCULATION PUMP to move the developer and
remove any air bubbles. Use the diagnostics to energize the PUMP.
LEVEL PROBES
Clean and check the PROBES.
The solution level in the
REPLENISHER TANK is low.
Mix new developer solution.
Solution does not flow through the
HOSES between the
REPLENISHMENT TANK and the
REPLENISHMENT PUMP.
Check
• HOSE CLAMPS are tight
• HOSES:
– are round and opened
– have no obstructions or air
bubbles
The DEVELOPER DRAIN VALVE
is open.
Close the VALVE.
SOLID STATE RELAY U2 that
controls the DEVELOPER
REPLENISHMENT PUMP
Check U2.
POPPET VALVES in the
DEVELOPER REPLENISHMENT
PUMP
Clean and check the VALVES.
DEVELOPER REPLENISHMENT
PUMP
Check FUSE F1 and the
REPLENISHMENT PUMP MOTOR B3.
The 5000 BOARD does not
energize the SOLID STATE
RELAY U2.
Check that the LED DS3 on the 5000
BOARD is illuminated. If necessary,
install a new BOARD.