Error Codes
5B6333 – 30SEP98
5
E011
Developer Heater failure
This error occurs when the 8000 CURRENT SENSE BOARD detects no
current through the DEVELOPER HEATER.
When the temperature of the
developer solution is too high, the
DEVELOPER HEATER
OVERTEMPERATURE
THERMOSTAT will open. The
THERMOSTAT will reset
automatically when the solution
cools.
Determine the cause for the high
developer temperature, and repair the
problem.
No continuity for the DEVELOPER
HEATER
The resistance should be
approximately 70 W.
Install a new HEATER.
SOLID STATE RELAY U1 that
controls the DEVELOPER
HEATER
Install a new U1.
The 5000 BOARD does not
energize the SOLID STATE
RELAY U1.
Check that LED DS4 on the 5000
BOARD is energized.If necessary,
install a new BOARD.
No AC power to the DEVELOPER
HEATER when both DS4 and DS9
are energized.
If the 5000 BOARD has removable
RELAYS, install a new RELAY K5004.
Otherwise, install a new 5000 BOARD.
8000 BOARD
Install a new BOARD.
E012
This error occurs when the 8000 CURRENT SENSE BOARD detects no
current through the FIXER HEATER.
When the temperature of the fixer
solution is too high, the FIXER
HEATER OVERTEMPERATURE
THERMOSTAT will open. The
THERMOSTAT will reset
automatically when the solution
cools.
Determine the cause for the high fixer
temperature and repair the problem.
No continuity for the FIXER
HEATER
The resistance should be
approximately 70 W.
Install a new HEATER.
SOLID STATE RELAY U5 that
controls the FIXER HEATER
Install a new U5.
The 5000 BOARD does not
energize the SOLID STATE
RELAY U5.
Check that the LED DS1 on the 5000
BOARD is energized. If necessary,
install a new BOARD.
No AC power to the FIXER
HEATER when both DS4 and DS9
are energized.
If the 5000 BOARD has removable
RELAYS, install a new RELAY K5004.
Otherwise, install a new 5000 BOARD.
8000 BOARD
Install a new BOARD.
E013
Level Sensor circuit failure
This error occurs when the circuit on the 5000 BOARD for the LEVEL
PROBE malfunctions. The following parts are disabled:
• DEVELOPER and FIXER REPLENISHMENT PUMPS
• RECIRCULATION PUMP
5000 BOARD
Install a new BOARD.
Code
Description
Possible Malfunction
Action