Internal Diagnostics
5B6333 – 30SEP98
23
Diagnostic Menus
Overview
When you enter the diagnostic menu, the software:
• de-energizes the following electrical components:
– all HEATERS
– all PUMPS
– all SOLENOIDS
– the BLOWER and DRIVE MOTORS
• de-energizes the SAFELIGHT OUTLET
• disables the error code detection
You can use the internal diagnostics to energize and de-energize electrical components. When you
energize a component, it will automatically de-energize in 4 minutes.
Details
To view the specific options that are available from the D1, D2, and D3 Menus, see the Flowcharts for
the Internal Diagnostics on the following pages.
Menus
D1 Menu
From the first menu, you can:
• access the History and Frequency Error Logs
• monitor the FILM DETECTOR SWITCHES
• monitor the SENSORS
• access the D2 Menu
• return to the main menu
VIEW
ERRORS
FILM
DETECT
SENSOR
TESTS
MORE
DONE/
RETURN