Proteus 1000 Operation Manual
49
Master Menu
MIDI Parameters
Front Panel Knob
Calibration
This function calibrates the minimum and maximum values for the five
front panel knobs. The controls should be calibrated periodically in order to
compensate for normal wear and tear on the controls themselves or if the
controls behave strangely. Before calibration, it’s usually a good idea to let
the unit warm up for about half an hour to ensure you have the maximum
range on all the front panel controls.
To Calibrate the Front Panel Control Knobs
1.
Press the Master menu button and advance to the Calibrate Knobs page.
2.
Press the Enter button to begin the calibration process. The following
screen appears.
3.
Turn all knobs to their minimum value (which is pegged in the counter-
clockwise direction), then press the Enter button. The screen briefly
flashes, “Reading Values…,” then displays:
4.
Turn all knobs to their maximum value (pegged in the clockwise
position), then press the Enter button. The display shows, “Reading
Values… Done.” Calibration is complete.
CALIBRATE KNOBS
Start
Turn all knobs to MIN
Press Enter when ready
Turn all knobs to MAX
Press Enter when ready