e-breathe Smartblower
Symbol
Meaning
Error
Action
Battery Warning
Capacity exhausted
• When switched on
20 sec.
• Permanent during
operation
• Acoustic alarm tone
• Optical alarm
• Acknowledgeable: Yes
The remaining battery run-
time is exhausted < 15 min
and/or the battery capacity
is < 5%.
At 3% / 0 min battery capa-
city the blower / SVE turns
off automatically for self-
protection.
On power-up:
Battery/SVE must be char-
ged.
During operation:
Work must be interrupted
and the battery / SVE must
be charged.
Battery Warning
Battery Communication
• When switched on
20 sec.
• During charging 10 sec.
• Permanent during
operation
• Acoustic alarm tone
• Optical alarm
• Acknowledgeable: Yes
The SVE unit‘s battery re-
gulator cannot commu-
nicate with the built-in
battery. As the battery is
malfunctioning or may be
deeply discharged.
On power-up/charging:
Battery/SVE must be char-
ged. If the battery is deeply
discharged, you can try to
recharge the battery by
charging it several times.
If the problem is not sol-
ved, have the blower tes-
ted by
e-breathe Service.
5.9 Alarms
The blower / SVE has the following warning devices to safely protect and inform the user.
- Acoustic alarm tone
- Optical alarm through flashing display
- Optical alarm by indication of the error on the display
Since, depending on the activity or working area, you may not be able to detect one of the warning alarms, all war-
ning alarms are always displayed simultaneously in the event of an alarm.
5.9.1 Warning Devices
5.9.2 Alarms
In the event of an alarm, work must be interrupted and the dangerous area must be left.
An alarm is signalled when the warning devices „audible alarm tone“ and „visual alarm“ are activated. An alarm always
sounds continuously or as long as the alarm reason/error still exists. Some alarms can be confirmed by briefly pressing
the menu button. The alarm will stop for 10 seconds and start again if the alarm cause still exists.
Alarms or errors that occur during the self check phase/on powering on are displayed for 20 seconds. The blower
automatically switches off after this period, so that operation is not possible without eliminating the fault.
In the event of an alarm, a skull with the corresponding error code and the alarm reason is displayed. If several alarms
occur at the same time, they are displayed alternately.