Pneumatic Lift
Options
Introduction
Activation and Operation
Introduction
Activation and Operation
Pneumatic Lift
The pneumatic lift accessory is a key piece of hardware to invest in if the flow
ratio test needs to be conducted. Also, if your material has a low flow rate and a
heavy weight is used, the pneumatic lift system lowers and raises the weight for
you.
Flow Ratio provides the user with insight into a materials shear thinning
ability and may correlate to molecular weight distribution. This test involves
obtaining flow rates at two different shear stresses, of generally a 10:1 ratio.
First, the lower mass weight is added to the piston. Next the higher mass weight
is added to the piston. The high mass weight may be well over 50 lbs. The
pneumatic lift automatically raises and lowers the weight safely.
You must enable the lift system in the SETUP mode, PNEU = ON. When
instrument is idle, the lift system is controlled by pressing the lift UP/ DOWN
arrows on the keypad.
There are also references to lift operation in the EDIT and SETUP modes.
Some notable settings are described:
Auto Purge =
At end of test, last flag or cut, drop weight to purge material.
Auto Raise =
If enabled, this option requests that the lift system raise the
weight
after
the experiment has ended
Auto Raise time =
If Auto Raise enabled, wait this long to raise the weight. See
following graph
Mini-Lift
This is a small Pneumatic Lift assembly that supports the piston rod itself. It
is useful for high-flow materials and can be used in conjunction with a melt
plug.
You must enable the Mini Lift system in the SETUP mode, MINI LIFT = ON.
When instrument is idle, the lift system is controlled by pressing SHIFT then the
lift UP / DOWN arrows on the keypad.
There are references to Mini Lift operation in the EDIT and SETUP modes.
Some notable settings are described:
MINI Lift Hold =
If enabled, this option requests that the lift system holds the
mini lift until the melt time has expired.
HOLD TIME =
If MINI Lift Hold enabled, wait this long to lower the MINI
Lift after raise time. Intrinsic Viscosity Option