Temperature (C)
LDPE
HDPE
Polybutene-1
polypropylene
120 0.797
--
0.806
0.880
130 0.791
--
0.800
0.872
140 0.785
--
0.794
0.864
150 0.780
0.780
0.787
0.852
160 0.777
0.777
0.780
0.840
170 0.770
0.770
0.774
0.819
180 0.765
0.765
0.767
0.758
190 0.760
0.760
0.760
0.754
200 0.755
0.755
0.754
0.750
210 0.748
0.748
0.746
0.746
220 0.744
0.744
0.740
0.742
230 0.738
0.738
0.733
0.738
240 0.733
0.733
0.726
0.734
250 0.737
0.727
0.719
0.730
Any inaccuracies in the melt density will propagated proportionally along to the
MFR values. Thus a 1% error in the melt density means a 1% accuracy error in
the MFR the value.
Programming Method B
This is a Method B setup for use with optical flags, (Small plastic strips with a
black area), if you have and encoder based machine skip this section go to the
"The Digital Encoder: Use and Calibration Chapter" Page 50.
Method B Flags Setup
1) Start a new program by pressing the EDIT SMART Key. -
The square cursor blinks in the upper left-hand corner of the screen prompting
the user to input the number of the program that will be created or edited.
2) Enter a number and ENTER on the numeric keypad.
Now you are editing a new program of number #. You can press the
EDIT
SMART Key to edit the existing program and select a B Method test
3) After pressing EDIT SMART Key, enter run parameters by pressing
EDIT again
The instrument is now in the VIEW Program mode starting with METHOD.
Press the
EDIT
SMART Key to enter run parameters for the method B
experiment. The 2nd line of the VFD display displays the parameter to be
edited. The 3rd line lists a group of suggested values. The bottom line of
SMART Keys is used to
ENTER, SELECT of SCROLL
data input,
respectively.
METHOD =
Press the
SMART Key
(browse)
until the 3rd line of the display has the B
over the
SELECT
SMART Key.
Press the
SELECT
SMART Key to select the method B test. The second line
should now read METHOD = B
Press SELECT/ENTER to accept the value and move to the SET POINT.
SETPOINT =
The user may enter a temperature on the numeric keypad and press enter or use
the
ENTER
,
SELECT
and
SCROLL
SMART Key keys to choose a typical
value.
Series 4000 Melt Indexer Operation Manual
Method "B" Run
33