Questions and
Answers
Typical Problems
PROBLEM
: Unit will not power up
SOLUTION
: Check the fuse in the input power module, where the power cable
connects to the instrument. The fuse is rated at 5 Amperes 120 Volts. Verify the
supply voltage is correct (e.g. 110 VAC vs. 220 VAC). If you have any
difficulty, please call Dynisco Polymer Test Technical Support.
PROBLEM
: The display works but the instruments keyboard does not respond.
SOLUTION
: The keyboards internal ribbon cable has become detached due to
rough handling during shipment. Remove the top cover, swing open front
display/keyboard panel mount. Locate keyboard ribbon cable coming through
bottom center of the display/keyboard panel mount. Reconnect to the right side
connector (J2) on the mother board. If you have any difficulty, please call
Dynisco Polymer Test Technical Support.
PROBLEM
: The digital temperature readout fluctuates, blinks or jumps
around:
SOLUTION
: The temperature of the barrel cannot physically change
temperature as fast as 1 degree C per second. If it does, usually a bad electrical
connection or failing electrical component can be blamed. Make sure the die
and piston rod are in the unit when heating. If you have any difficulty, please
call Dynisco Polymer Test Technical Support.
PROBLEM
: The VFD on the instrument displays a negative temperature.
SOLUTION
: The RTD is in a shorting condition. Perform an RTD Display test.
If the reading is abnormal call Dynisco Polymer Test Technical Support.
PROBLEM
: Temperature is stable but not reaching set point.
SOLUTION
: Heat the instrument with the die and piston rod assembly in the
sample chamber. If you have any difficulty, please call Dynisco Polymer Test
Technical Support.
PROBLEM
: The keyboard works but the instruments display does not work.
SOLUTION
: The VFD's internal ribbon cable has become detached due to
rough handling during shipment. Remove the top cover, swing open front
display/keyboard panel mount. Locate VFD ribbon cable. Reconnect to the left
side connector (J1) on the mother board. Check connection to the VFD itself. If
you have any difficulty, please call Dynisco Polymer Test Technical Support.
Series 4000 Melt Indexer Operation Manual
Questions and Answers
69