5-
1
Chapter 5 Replacement Parts
5.1 Consumables and Spare Parts
The following parts are consumables. Contact the nearest Dynamica agent before
consuming all of them.
WI Lamp (Tungsten Iodide Lamp): P/N: 25-03-0001
D
2
Lamp (Deuterium Lamp): P/N: 25-02-0013
Main Power Fuse: P/N: 13-02-0004
Table 5-1 Guaranteed life of lamp
Guaranteed
time
Guaranteed Period
WI Lamp
1000h
6 months after delivery to customer’s site
D
2
Lamp
1000h
6 months after delivery to customer’s site
5.2 Part Replacement
5.2.1 Replacement of Light Source Lamp
The following shows how to replace the existing lamp with a new one.
(1) Turn off the power switch. Pull off the power cord and wait for at least 20
minutes until the lamp cools down.
(2) Remove the fastening screw from the side face of the light source cover,
and then remove the cover. (See Fig. 5-1)
Cautions on Replacement of Lamp
Do not replace the lamp with the power switch on. Note that high voltage is applied
to the lamp.
High voltage is discharged while the power is turned off.
Wait for at least 10 minutes after the power switch is turned on.
Wear clean gloves for mounting a new lamp. Do not leave fingerprints on the lamp
tube wall.
For D
2
lamp, care should be taken to prevent leaving fingerprints on the protruded
sections.