11/12/02
15:38
600.0
nm
0.854
ABS
Multipl WL
1
Parameter Setup
1
Data Mode
ABS
2
Sample Setup
2
NUM WL
2
3
Curve Setup
3
WL Setup
500.0
4
Print Setup
4
Baseline
System
5
System Setup
6
Save Parameter
7
User Baseline
0
Forward
Input Value:_
(2 to 6)
Specifies the number of wavelengths.
The following shows the relation between the number
of wavelengths and each data mode.
Data
Mode
Settable
Wavelengt
h
Description
T%
2 to 6
Transmittance at each
wavelength is measured
and displayed.
ABS
2 to 6
Absorbance at each
wavelength is measured
and displayed.
2 2–wavelength
calculation
is made and concentration
is obtained.
WL
Conc
3 3–wavelength
calculation
is made and concentration
is obtained.
(a) WL NUM (Data Mode: Conc)
The photometric values A (WL1) and A (WL2) at two specified wavelength
is calculated based for the following formula.
D = WL2 – WL1
3-48