Press <5> to select Parameter Setup from the time scan submenu, and a
submenu list will appear on the right side of the screen as shown in the
illustration below:
11/12/2002
14:36
600.0
nm
0.854
ABS
Time Scan
1
Parameter Setup
2
Sample Setup
3
Print Setup
4
System Setup
5
Save Parameter
0
Forward
Input File No._
(1 to 20)
Use the keypad to select the desired catalog number (1-20), then press ENTER.
If the memory location that you have selected is available, the instrument stores
entire variable set under test name you have specified. If the catalog number you
requested is already full, the following message appears:
Overwrite this File?
To delete the previous test variables and store the new variable set, press
ENTER. To leave the old variables set intact and store new variables in a
different location, press RETURN.
6. Forward:
End setting up and go to the measurement display.
3-45