1-2
1.2 Installation site
1.2.1 Installation conditions
Power requirements:
Power supply voltage: AC100, 115, 220, 230 or 240V (There must be the
devices for switching voltage.)
Allowable fluctuation is within ±10% of rated voltage.
Frequency: 50 or 60 Hz
Allowable fluctuation is within ±4% of rated frequency.
Power capacity:
200VA or more.
Prepare a capacity more than 500VA in reference with
a co-use with other accessory devices.
Note: If warning labels are not readable because of stains on coming off,
please notify local Dynamica sales representative or service office.
New labels will be provided.
Grounding line:
Grounding line resistance of 100
Ω
or less is required.
Place for Installation
Minimum floor area:
Floor area with a width of 550 mm and a depth of 710
mm or more is required.
Floor strength:
Floor strength 30kg or more.
Select a rigid and horizontal plane capable of
withstanding the above load as an installation area.
1.2.2 Installation Environment
Operating temperature: 5~35
℃
.
In order to perform a measurement under the most stable condition, we
recommend that the instrument is used in an air conditioning room of 20
℃
~ 25
℃
.
Operating humidity: 45% ~ 85%.