11/04/02
10:17
600.0
nm
0.854
ABS
Wavelength Scan
1
Parameter Setup
1
Printer Type
ESC/P
2
Sample Setup
2
Auto Print
ON
3
Print Setup
3
Graph Print
ON
4
System Setup
4
Data Print
Peak
5
Save Parameter
5
List Interval (nm)
10.0
6
User Baseline
0
Forward
Select Function: _
Printer Type
There are three kinds of printer to select: EPSON, Seiko, HP.
Auto Print
If the autoprint function is “ON”, the DB-20 will print automatically the setting
parameter and the graph or data after completion of a measurement. But if the
autoprint function is “OFF”, the instrument will not print automatically unless
the Print key is pressed.
Graph Print
Using this parameter to determine whether to output the on-screen spectral
graph onto the printer automatically at the end of wavelength scanning.
Data Print
If it is desired to print out photometric values at a fixed wavelength interval
during spectral scanning, set the Text Print parameter to Peak, Valley or All
Data. Otherwise, set this parameter to Not Print.
List Interval (nm)
If the Text Print parameter is set to ON, specify a wavelength interval at which
photometric values are printed out.
This parameter is specifiable within a range of 0.1 to 100.0 nm in increments
of 0.1 nm.( Please refer to section 3.4.4 )
4. System Setup
Press <4> to select system setup from the Wavelength Sub Menu and a sub menu
list will appears on the right side of the screen as shown in the illustration below.
These parameters are the same as 3.3 except Peak Threshold.
3-24