20
Dymax BlueWave® LED Flood System User Guide
3.
Press and release the
Button to enter the Options Menu (Figure 21).
4.
Use
or
Arrows so that “ADJUST TIMER” is flashing
and press the
Button.
5.
The Adjust Timer Screen (Figure 22) will display. Use the Directional Keys to adjust the Timer to the
correct value. The Timer can be adjusted from 0000.0 seconds to 9999.9 seconds. Press and release the
Button when you have entered the desired activation time.
Figure 20. Time Mode
Main Screen
Figure 21. Options Menu
Figure 22. Adjust Timer
Screen
6.
Press the Foot Switch or the
Button on the front panel to start the Timer. An exposure cycle can
be cancelled at any time by momentarily pressing the
Button or Foot Switch.
Operating in PLC Mode
PLC Mode Description
CAUTION!
Always wear protective goggles or a face shield when working near UV light. The
PLC is capable of turning on the LED at any time, or when a user selects PLC Mode from the front
panel controls.
PRÉCAUTIONS!
Veuillez toujours porter des lunettes protectrices pour les yeux ou pour le
visage lors de travail en contact avec la lumière UV. Le 'PLC' peut mettre sous tension le système
DEL en tout temps, ou lorsque l'utilisateur sélectionne le mode 'PLC' avec les commandes du
panneau frontal.
PLC Mode
means that an external PLC (Programmable Logic Controller) controls the system. The PLC provides
input signals to the flood unit to control the LED. The PLC monitors the status of the flood unit by reading
output signals provided by the unit. The PLC is typically programmed to start other machinery when the flood
unit becomes ready, or perhaps start the flood unit when other machinery is ready. The input signals and
output signals are present on a connector on the back of the power supply labeled
‘PLC’
.
NOTE
: Output signals are always provided in manual mode and timer mode to allow simple interfacing to
alarms, counters, other interlocks, etc. even if a PLC will not be controlling the unit. However, the input signals
will be ignored in manual mode and timer mode.