Dymax BlueWave® LED Flood System User Guide
Back Panel
The back panel includes these components and connection points:
■
Power Cord Receptacle
—
Connection point for the power cord.
■
I/O (On/Off) Switch
—
Moving the switch to the I (on) position powers up the power supply. Moving the
switch to the O (off) position cuts power to the power supply.
■
Cooling Air Exhaust
—
Cooling air enters the side vents and exhausts out the rear of the unit.
NOTE:
The exhaust vents on the rear of the power supply must not be blocked or the unit may overheat
and shut down.
Ensure 1" [2.54 cm] of clearance is provided around the power supply and at least 6"
[15.24 cm] of clearance is provided around the cooling fan inlets and outlet.
■
Foot-Switch Jack
—
Connection point for the foot switch. Pressing the foot switch starts a curing cycle.
In timed operation, pressing and releasing the foot switch activates the curing array for the length of time
displayed in the cycle time display. A second press will terminate a timed cure cycle immediately. In
manual operation, pressing and holding the foot switch activates the curing light until the foot switch is
released.
■
PLC Access Switch
–
Button used to temporarily unlock the option to access PLC mode.
■
PLC Connector
—
Connection point for a user-supplied cable to connect the unit to a PLC for remote
operation. Connection is a standard 25-pin D-style connector. Refer to page 19 for details on the
connections to this port.
■
Array Cable Connector
—
Connection point for the interconnection cable connecting the power supply
to the array.
Figure 3. Back Panel
Intensity Control Feature
The components used in all light-curing systems degrade with use. Therefore, the maximum intensity
decreases as exposure hours accumulate. Setting process intensity requirements lower than the maximum
enables the
BlueWave LED Flood
’s
intensity control feature to allow for compensation of gradual decreases in
light intensity.
The unit's intensity can be adjusted using the intensity adjustment on the front control panel as a percentage
of full scale intensity.