![Durr Dental VistaScan Mini View Installation And Operating Instructions Manual Download Page 29](http://html.mh-extra.com/html/durr-dental/vistascan-mini-view/vistascan-mini-view_installation-and-operating-instructions-manual_2542936029.webp)
2142100002L02 1908V007
27
Assembly
Testing the device
You can scan in an X-ray image to check that
the unit is properly connected.
i
Select the
Test
tab.
i
Select the unit from the
Registered Units
list.
i
Select the mode class.
i
Select the mode.
i
Click on
Scan Image
.
i
Scan an image plate, see "11.2 Scanning the
image data via a computer".
i
Activate the connected unit in the
Registered
column.
You can also register multiple units.
The
VistaNet device configuration
window al-
lows you to change the device name (
name
),
manually enter an IP address or call up informa-
tion.
i
Click
.
i
If necessary change the
name
.
i
Click
Apply
to save the configuration.
Entering a fixed IP address (recommended)
To reset the network settings, keep the
unit reset key pressed for 15 - 20 sec-
onds while switching on.
i
Deactivate
DHCP
.
i
Enter the IP address, subnet mask and gate-
way.
i
Click on
Apply
.
The configuration is saved.
EN