2142100002L02 1908V007
25
Assembly
8 Commissioning
NOTICE
Short circuit due to the build up of
condensation
i
Do not switch on the unit until it has
warmed up to room temperature and
it is dry.
8.1 Installing and configuring the
unit
The unit supports the following imaging pro-
grams:
– VistaSoft from Dürr Dental
– VistaConnect from Dürr Dental
– DBSWIN from Dürr Dental
– VistaEasy from Dürr Dental
– Third-party software on request
Configuring the network
Network configuration
Various options are available for network config-
uration:
– Automatic configuration via DHCP.
– Automatic configuration via Auto-IP for direct
connection of unit and computer.
– Manual configuration.
i
Configure the network settings of the unit us-
ing the software or, if applicable, the touch
screen.
i
Check the firewall and release the ports, if ap-
plicable.
Network protocols and ports
Port
Purpose
Serv-
ice
45123 UDP,
45124 UDP
Unit recognition and
configuration
2006 TCP
Unit data
514
1)
UDP
Event protocol data
Syslog
2005 TCP,
23 TCP
Diagnosis
Telnet,
SSH
1)
The port can vary depending on the config-
uration.
When the unit is first connected to a
computer, it applies the language and
time settings of the computer.
i
If it is not 100% clear from the unit data sheet
that such connections can be safely made or
if you are in any doubt, always get a suitably
qualified person (e.g. the manufacturer) to ver-
ify that the setup is safe.
i
Observe the specifications of IEC 60601-1
(EN 60601-1) when connecting the appliance
with other appliances, e.g. a PC system, both
in and outside the patient environment.
i
Only connect peripheral units (such as com-
puters, monitors or printers) that conform at
least to the requirements set out in
IEC 60950-1 (EN 60950-1).
i
The connected computer must conform to
EN 55032 (class B) and EN 55024.
A copy of the system manufacturer‘s
declaration in accordance with Article 12
of Directive 93/42/EEC can be found in
our download section at www.duerrden-
tal.com (document no. 9000-461-264).
Connecting the unit via the network cable
i
Remove the cover from the rear of the device.
i
Connect the supplied network cable to the
network connection of the device.
i
Refit the cover.
The cover on the rear must be correctly
fitted when the device is operated within
the patient environment.
EN