40
2142100002L02 1908V007
Usage
Use the following cleaning and disinfecting im-
mersion disinfectants:
– ID 213 Instrument disinfection
– ID 212
– ID 212 forte
NOTICE
Heat can damage plastic parts.
i
Do not use a thermal disinfector or
steam steriliser on any parts of the
device.
i
Touch
on the touch screen.
The fixing mechanism moves into the clean-
ing position.
i
Press the release button and remove the cov-
er upwards at the same time.
2
1
i
Remove the fixing mechanism by moving it
upwards.
12 Cleaning and disinfection
When cleaning and disinfecting the unit and its
accessories, observe country-specific directives,
standards and specifications for medical prod-
ucts as well as the specific specifications for
dental practices and clinics.
NOTICE
The use of unsuitable agents and
methods can damage the unit and
accessories.
Do not use any products based on phe-
nolic compounds, halogen-releasing
compounds, strong organic acids or ox-
ygen-releasing compounds, as they
may damage the materials.
i
Dürr Dental recommends using disin-
fectants from the Dürr Dental product
range. Only the products specified in
these instructions have been subject-
ed to material compatibility testing by
Dürr Dental.
i
Read the operating instructions for
the disinfectants.
Wear protective gloves.
12.1 Image plate scanner
Unit surfaces
The unit surface must be cleaned and disinfect-
ed of any contamination or visible soiling.
Dürr Dental recommends using the disinfectants
FD 322, FD 333, FD 350 and FD 366 sensitive.
NOTICE
Liquid can cause damage to the unit.
i
Do not spray the unit with cleaning
and disinfectant agents.
i
Make sure that liquid does not get in-
side the unit.
i
Remove any soiling with a soft, lint-free cloth
that has been dampened with cold tap water.
i
Disinfect the surfaces using a disinfection
wipe. Alternatively, use disinfectant on a soft,
lint-free cloth.
Input unit
The input unit must be cleaned and disinfected
if there are indications of contamination or visi-
ble dirt.
EN