4 | Product description
D-R 290
49
Equation
7
Table 4.10: Equation 7
The dust concentration
c is:
Equation
8
Table 4.11: Equation 8
I
0
emitted light
k
extinction coefficient
I
received light
OD Optical density
l
Measurement path length, (x 2 for
auto-collimation)
c
Dust concentration
b
Dust coefficient
a
Dust offset
For the reasons already mentioned, the coefficients
a and b must be determined by
gravimetric measurement. The measurements to be performed must be performed
with the occurring plant loads and the difference filtering possibilities of the respective
plant. Check measurements for different types of fuel are necessary. Only when these
comparisons figures are available can the extinction values be correctly evaluated re
garding the dust emission. Since the dust concentration is subject to statistical fluctua
tions, statistical methods are best for determining the calibration curve for the relation
ship between
and dust concentration.
For further information please refer to the standards VDI 3950 (Germany), EN 14181,
EN 13284-1, EN 15259 and others.
The setting of the coefficients in the device is described in section
.
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...